XEN 45 GTS
Report
- Report Number
- 3011299751-2023-00173
- Event Type
- Injury
- Date Received
- November 3, 2023
- Date of Event
- October 19, 2023
- Report Date
- November 3, 2023
- Manufacturer
- ALLERGAN (IRVINE)
- Product Code
- KYF
- UDI-DI
- 01240000000405
- PMA / PMN Number
- K161457
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
CONTINUED H.6. HEALTH EFFECT - IMPACT CODES: F22. ARTICLE CITATION: ZHANG, YAO, ET AL. ¿RECANALIZATION OF XEN45 GEL STENT IMPLANT OCCLUSION USING 10 - 0 NYLON SUTURE IN REFRACTORY GLAUCOMA: A CASE REPORT.¿ BMC OPHTHALMOLOGY, VOL. 23, NO. 1, HTTPS://DOI.ORG/10.1186/S12886-023-03109-7. FURTHER INFORMATION REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. THE EVENT IS A KNOWN POTENTIAL ADVERSE EVENT ADDRESSED IN THE PRODUCT LABELING.
REPORTED EVENTS OF "PATIENT UNDERWENT XEN®45 GTS AB-INTERNO IMPLANTATION IN THE RIGHT EYE. POD1, IOP WAS 5 MMHG WITH A CHOROIDAL DETACHMENT. POD5, IOP WAS 10 MMHG WITH A DIFFUSE BLEB AND CHOROIDAL DETACHMENT HAD RESOLVED. POD6, IOP WAS 40 MMHG AND MATERIAL OCCLUDING THE INTERNAL OSTIUM OF THE STENT. POD12, ND:YAG LASER SHOCKWAVE THERAPY WAS PERFORMED, RESULTING IN AN IOP DECREASE FROM 42 MMHG TO 25 MMHG AND OCCLUSION RESOLVED. THEN 6 HOURS LATER, IOP INCREASED TO 40 MMHG AND MULTIPLE ROUNDS OF SHOCKWAVE THERAPY WITH INCREASING INTENSITY WERE PERFORMED, BUT OCCLUSION REMAINED. IOP REMAINED SIGNIFICANTLY ELEVATED. POD14, REVISION SURGERY WAS PERFORMED VIA CONJUNCTIVAL INCISION WITH FIBROTIC REMOVAL, BUT NO FLOW WAS OBSERVED; FURTHER RECANALIZATION WAS PERFORMED VIA A 10-0 NYLON SUTURE TO THREAD THE STENT RESULTING IN IMMEDIATE FLOW OF AQUEOUS HUMOR AND RESOLVING THE OCCLUSION. POD15, IOP WAS 8 MMHG WITH A FILTERING BLEB. POD21, IOP WAS 10 MMHG, AND REMAINED AT 12 MMHG THROUGH POM6. AT POM11.5, IOP WAS 15 MMHG WITHOUT MEDICATION." WERE NOTED IN THE ARTICLE: "RECANALIZATION OF XEN45 GEL STENT IMPLANT OCCLUSION USING 10¿-¿0 NYLON SUTURE IN REFRACTORY GLAUCOMA: A CASE REPORT." BMC OPHTHALMOL. 2023 OCT 19;23(1):418.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2221577 | XEN 45 GTS | IMPLANT, EYE VALVE | KYF | ALLERGAN (IRVINE) | NI | 01240000000405 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 16 YR | Female | Required Intervention |