FDA Adverse Event Injury Summary report: N

BALL HEADS: MECTACER BIOLOX DELTA CERAMIC BALL HEAD 12/14 Ø 36 SIZE S

MDR report key: 18063342 · Received November 3, 2023

Report

Report Number
3005180920-2023-00851
Event Type
Injury
Date Received
November 3, 2023
Date of Event
October 11, 2023
Report Date
November 3, 2023
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LZO
UDI-DI
07630030809194
PMA / PMN Number
K112115
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

BATCH REVIEW PERFORMED ON 12 OCTOBER 2023 LOT 2303248: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 08-MARC-2023. EXPIRATION DATE: 2028-02-16. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. ADDITIONAL COMPONENT INVOLVED IN THE EVENT: LINER: MPACT 01.32.3648HCT FLAT PE HC LINER Ø36/F (K103721) LOT 2242600: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 14-DEC-2022. EXPIRATION DATE: 2027-11-29. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW.

Description of Event or Problem · 0

AT ABOUT 3 MONTHS FROM THE PRIMARY, THE PATIENT CAME IN REPORTING PAIN DUE TO A DISLOCATION OF THE HEAD FROM THE LINER. THE SURGEON REVISED THE HEAD AND LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2149278 BALL HEADS: MECTACER BIOLOX DELTA CERAMIC BALL HEAD 12/14 Ø 36 SIZE S HIP FEMORAL HEAD LZO MEDACTA INTERNATIONAL SA 2303248 07630030809194

Patients

Seq Age Sex Outcome Treatment
1 Unknown Required Intervention