(01)00607915125318(17)260626(30)01(10)1495840
Report
- Report Number
- 2027111-2023-00696
- Event Type
- Malfunction
- Date Received
- November 2, 2023
- Date of Event
- October 23, 2023
- Report Date
- May 24, 2024
- Manufacturer
- APPLIED MEDICAL RESOURCES
- Product Code
- FZP
- UDI-DI
- 00607915125318
- PMA / PMN Number
- K011236
- Removal / Correction Number
- 2027111-01/26/24-001-R
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- 003
Narratives
THE EVENT UNIT IS ANTICIPATED TO RETURN TO APPLIED MEDICAL. A FOLLOW UP REPORT WILL BE PROVIDED FOLLOWING THE COMPLETION OF THE INVESTIGATION.
APPLIED MEDICAL HAS ISSUED A VOLUNTARY RECALL OF OUR EPIX UNIVERSAL CLIP APPLIER FOR SPECIFIC CA500 LOTS. THESE CA500 UNITS ARE BEING RECALLED DUE TO A NONCONFORMANCE THAT MAY RESULT IN A CLIP NOT LOADING INTO THE JAWS AFTER THE TRIGGER IS ACTUATED. APPLIED MEDICAL HAS RECENTLY IMPLEMENTED MANUFACTURING CORRECTIONS WHICH ARE INTENDED TO REDUCE THE POTENTIAL FOR THIS TYPE OF INCIDENT TO REOCCUR. THE LOT ASSOCIATED WITH THIS COMPLAINT, #(B)(4), WAS INCLUDED IN THE RECALL.
PROCEDURE PERFORMED: LAPAROSCOPIC CHOLECYSTECTOMY EVENT DESCRIPTION: TWO CA500 LOT #1495840 COMPLAINTS OCCURRED DURING THE SAME PROCEDURE: (B)(4). BOTH DEVICES MISFIRED. THERE WERE "LOTS OF TENSION" AND THE CLIPS WERE NOT CLOSING COMPLETELY. THE ISSUE WAS INITIALLY OBSERVED WHEN THE FIRST CLIP WAS LOADED. THE SURGEON TRIED TO FIRE ABOUT THREE TIMES WITH NO SUCCESS BEFORE OPENING ANOTHER DEVICE. THE SAME ISSUE OCCURRED WITH THE SECOND DEVICE. THE SURGEON OPENED UP A THIRD CA500 DEVICE TO COMPLETE THE CASE. A 5 MM ADVANCED FIXATION TROCAR WAS USED. A CLIP SAT PROPERLY INTO THE JAWS. NO PRESSURE WAS APPLIED TO THE TRIGGER WHILE MOVING THROUGH THE TROCAR. NO CLIP WAS LOADED INTO THE JAWS PRIOR TO THE DEVICE¿S INSERTION/REMOVAL THROUGH THE TROCAR. NO CLIP WAS MANUALLY REMOVED AS A LOADED CLIP. THE TRIGGER APPEARED INTACT. THE CLIP FULLY LOADED INTO THE JAWS UPON ACTUATION. THE TRIGGER WAS SQUEEZED ¿PLASTIC TO PLASTIC". THE SURGEON FULLY SKELETONIZED THE VESSEL PRIOR TO USING THE CLIP APPLIER. THE SURGEON DID NOT USE THE CLIP APPLIER TO SKELETONIZE TISSUE. THE APEX OF THE CLIP WOULD NOT CLOSE. PHOTOS ARE AVAILABLE. NO PATIENT INJURY. PRODUCTS ARE AVAILABLE FOR RETURN. PATIENT STATUS: NO PATIENT INJURY. INTERVENTION: THE SURGEON OPENED UP A THIRD CA500 DEVICE TO COMPLETE THE CASE.
PROCEDURE PERFORMED: LAPAROSCOPIC CHOLECYSTECTOMY. EVENT DESCRIPTION: TWO CA500 LOT #1495840 COMPLAINTS OCCURRED DURING THE SAME PROCEDURE: COMPLAINT 1 OF 2: (B)(4). COMPLAINT 2 OF 2: (B)(4). BOTH DEVICES MISFIRED. THERE WERE "LOTS OF TENSION" AND THE CLIPS WERE NOT CLOSING COMPLETELY. THE ISSUE WAS INITIALLY OBSERVED WHEN THE FIRST CLIP WAS LOADED. THE SURGEON TRIED TO FIRE ABOUT THREE TIMES WITH NO SUCCESS BEFORE OPENING ANOTHER DEVICE. THE SAME ISSUE OCCURRED WITH THE SECOND DEVICE. THE SURGEON OPENED UP A THIRD CA500 DEVICE TO COMPLETE THE CASE. A 5 MM ADVANCED FIXATION TROCAR WAS USED. A CLIP SAT PROPERLY INTO THE JAWS. NO PRESSURE WAS APPLIED TO THE TRIGGER WHILE MOVING THROUGH THE TROCAR. NO CLIP WAS LOADED INTO THE JAWS PRIOR TO THE DEVICE¿S INSERTION/REMOVAL THROUGH THE TROCAR. NO CLIP WAS MANUALLY REMOVED AS A LOADED CLIP. THE TRIGGER APPEARED INTACT. THE CLIP FULLY LOADED INTO THE JAWS UPON ACTUATION. THE TRIGGER WAS SQUEEZED ¿PLASTIC TO PLASTIC". THE SURGEON FULLY SKELETONIZED THE VESSEL PRIOR TO USING THE CLIP APPLIER. THE SURGEON DID NOT USE THE CLIP APPLIER TO SKELETONIZE TISSUE. THE APEX OF THE CLIP WOULD NOT CLOSE. PHOTOS ARE AVAILABLE. NO PATIENT INJURY. PRODUCTS ARE AVAILABLE FOR RETURN. PATIENT STATUS: NO PATIENT INJURY. INTERVENTION: THE SURGEON OPENED UP A THIRD CA500 DEVICE TO COMPLETE THE CASE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2111988 | (01)00607915125318(17)260626(30)01(10)1495840 | CLIPS, IMPLANTABLE | FZP | APPLIED MEDICAL RESOURCES | CA500 | 1495840 | 00607915125318 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | 5 MM ADVANCED FIXATION TROCAR |