PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM
Report
- Report Number
- 2024168-2023-12152
- Event Type
- Injury
- Date Received
- November 1, 2023
- Date of Event
- October 4, 2023
- Report Date
- January 12, 2024
- Manufacturer
- ABBOTT VASCULAR
- Product Code
- MGB
- PMA / PMN Number
- P960043
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
MANUFACTURER'S INVESTIGATION IS STILL PENDING AT THIS TIME. RESULTS AND CONCLUSIONS WILL BE PROVIDED IN THE FINAL REPORT.
THE DEVICE WAS RETURNED FOR ANALYSIS. THE REPORTED CUFF MISS WAS NOT CONFIRMED AS NOT ALL COMPONENTS WERE RETURNED FOR ANALYSIS. A REVIEW OF THE MANUFACTURING RECORDS IDENTIFIED NO MANUFACTURING NONCONFORMITIES ISSUED TO THE REPORTED LOT THAT WOULD HAVE CONTRIBUTED TO THIS EVENT. THE REPORTED DIFFICULTY AND SUBSEQUENT TREATMENT APPEAR TO BE RELATED TO AN INTERACTION OF THE DEVICE WITH PATIENT ANATOMY OR INABILITY TO MAINTAIN POSITION/STABILITY OF THE DEVICE DURING DEPLOYMENT DUE TO CIRCUMSTANCES OF THE PROCEDURE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO THE DESIGN, MANUFACTURE, OR LABELING OF THE DEVICE.D4 - LOT NUMBER UPDATED FROM 2101041 TO 2110442. D4: EXPIRATION DATE AND H4: MANUFACTURING DATE CORRECTED AS A RESULT.
IT WAS REPORTED THAT AN ARTERIOTOMY CLOSURE OF THE RIGHT COMMON FEMORAL ARTERY WAS ATTEMPTED WITH A PROGLIDE DEVICE AFTER A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY INTERVENTIONAL PROCEDURE USING A 6F SHEATH. REPORTEDLY, THE SUTURE WAS NOT PRESENT WHEN THE PLUNGER WAS REMOVED (A CUFF MISS OCCURRED) AFTER STEP 3. A NEW PROGLIDE DEVICE WAS USED TO ACHIEVE HEMOSTASIS. THERE WAS NO ADVERSE PATIENT SEQUELA AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2142333 | PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM | SUTURE MEDIATED CLOSURE | MGB | ABBOTT VASCULAR | 2110442 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Required Intervention |