Narratives

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # (B)(4). D4: THE DEVICE CATALOG NUMBER IS UNKNOWN; THEREFORE, UDI IS UNAVAILABLE. PRODUCT CODE 113785045, LOT J9636C WAS REPORTED BY THE INITIAL REPORTER. FOLLOW-UP IS BEING CONDUCTED IN ORDER TO DETERMINE THE ASSOCIATED REGULATORY INFORMATION. A SUPPLEMENTAL REPORT WILL BE SUBMITTED IF ADDITIONAL INFORMATION IS RECEIVED THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803 (AND/OR PART 4, AS APPLICABLE). THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # (B)(4). INVESTIGATION SUMMARY: ACCORDING TO THE INFORMATION PROVIDED, "IT WAS REPORTED THAT THIS WAS A TSA FOR NA PERFORMED ON (B)(6) 2023. AT A TRIAL, A PEG SIZE 44 MM IMPLANT TRIAL FITTED THE AFFECTED SITE, BUT THE REAL PEG SIZE 44 MM IMPLANT WAS UNABLE TO BE INSERTED. THEREFORE, THE SURGEON EXCHANGED IT TO A PEG SIZE 40 MM IMPLANT AND INSERTED IT. THE SURGERY WAS COMPLETED SUCCESSFULLY WITH OVER 30 MINUTES. THERE WAS NO HARM TO THE PATIENT. NO FURTHER INFORMATION IS AVAILABLE.". THE GLB SHD END PEG GLEN 44 GBB (P/N: 113785045, LOT # J9636C) DEVICE ASSOCIATED WITH THIS REPORT WAS RETURNED TO DEPUY SYNTHES FOR EVALUATION. VISUAL ANALYSIS FOUND DEEP SCRATCHES ON THE ARTICULATING SURFACE OF THE PEGGED GLENOID AS WELL AS GOUGES FOUND ON THE INFERIOR EDGE OF GLENOID BONE INTERFACE SURFACE. THE DAMAGE IS CONSISTENT WITH EXTRACTION ATTEMPTS AFTER INSERTION. THE POLY SURFACE INCLUDING ALL FIVE PEGS ARE FREE OF ANY SURFACE LEVEL DEFECTS/BURRS THAT COULD INHIBIT THE DEVICE FROM BEING INSERTED. THERE IS NO EVIDENCE TO SUGGEST THE DEVICE WAS UNABLE TO BE INSERTED INTRAOPERATIVELY. DIMENSIONAL ANALYSIS OF THE OVERALL DEVICE AND ALL FIVE PEGS WAS PERFORMED. PER DRAWING DWG-113785045 REV. B, ALL DIMENSIONS MEASURED OF THE DEVICE WERE FOUND TO BE WITHIN TOLERANCE SPECIFICATIONS (SEE ATTACHMENT FOR MEASUREMENT DETAILS). THE OVERALL COMPLAINT COULD NOT BE CONFIRMED BASED ON THE EVIDENCE AND INFORMATION AVAILABLE. THE DAMAGE ON THE SURFACE OF THE DEVICE IS ATTRIBUTED TO USER EXTRACTION INTRAOPERATIVELY. NO FURTHER CONCLUSION CAN BE DRAWN. BASED ON THE INVESTIGATION FINDINGS, NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE COMPLAINT CONDITION. AS PART OF DEPUY SYNTHES QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE HISTORY LOT: A SEARCH OF THE DEPUY SYNTHES QUALITY SYSTEM WAS PERFORMED FOR THE FINISHED DEVICE LOT NUMBER J9636C, AND NO NON-CONFORMANCES OR MANUFACTURING IRREGULARITIES WERE IDENTIFIED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THIS WAS A TSA FOR NA PERFORMED ON (B)(6) 2023. AT A TRIAL, A PEG SIZE 44 MM IMPLANT TRIAL FITTED THE AFFECTED SITE, BUT THE REAL PEG SIZE 44 MM IMPLANT WAS UNABLE TO BE INSERTED. THEREFORE, THE SURGEON EXCHANGED IT TO A PEG SIZE 40 MM IMPLANT AND INSERTED IT. THE SURGERY WAS COMPLETED SUCCESSFULLY WITH OVER 30 MINUTES. THERE WAS NO HARM TO THE PATIENT. NO FURTHER INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2226334 GLB SHD END PEG GLEN 44 GBB SHOULDER GLENOID  KWT DEPUY ORTHOPAEDICS INC US J9636C

Patients

Seq Age Sex Outcome Treatment
1 Unknown