FDA Adverse Event Malfunction Summary report: N

TELESCOPE, 10 MM, 0°, QUICK LOCK, AUTOCLAVABLE

MDR report key: 18040016 · Received October 31, 2023

Report

Report Number
9610773-2023-03119
Event Type
Malfunction
Date Received
October 31, 2023
Date of Event
October 4, 2023
Report Date
August 12, 2024
Manufacturer
OLYMPUS WINTER & IBE GMBH
Product Code
HET
UDI-DI
04042761001557
PMA / PMN Number
K923982
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

G4:K923982/ K912362. THE SUBJECT DEVICE WAS RETURNED TO AN OLYMPUS REPAIR CENTER FOR EVALUATION AND THE REPORTED ISSUE WAS CONFIRMED. THE DEVICE EVALUATION FOUND THAT THE LIGHT GUIDE (LG) POST WAS BROKEN OFF. IT WAS ALSO FOUND THAT THERE WERE MINOR DENTS ON THE OUTER TUBE, DISTAL FIBER BREAKAGE AND DEBRIS IN OPTICAL SYSTEM. THE INVESTIGATION IS ONGOING. A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION OR IF ANY ADDITIONAL INFORMATION IS PROVIDED.

Additional Manufacturer Narrative · 0

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE THE RESULTS OF THE LEGAL MANUFACTURER'S FINAL INVESTIGATION. A REVIEW OF THE DEVICE HISTORY RECORD FOUND NO DEVIATIONS THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE. IT HAS BEEN OVER 23 YEARS SINCE THE SUBJECT DEVICE WAS MANUFACTURED. THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION, AND THE CUSTOMER'S REPORTABLE MALFUNCTION WAS CONFIRMED. A DEFINITIVE ROOT CAUSE WAS NOT IDENTIFIED, HOWEVER BASED ON THE RESULTS OF THE INVESTIGATION, THE PROBABLE CAUSE OF THE MALFUNCTION WOULD LIKELY BE DUE TO EXCESSIVE USE AND, DUE TO THE AGE OF THE DEVICE, WEAR AND TEAR. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Description of Event or Problem · 0

THE CUSTOMER REPORTED TO OLYMPUS, THE TELESCOPE, 10 MM, 0°, QUICK LOCK, AUTOCLAVABLE HAD A BROKEN LIGHT POST. THERE WERE NO REPORTS OF PATIENT HARM.