4003409
Report
- Report Number
- 1218950-2023-00830
- Event Type
- Malfunction
- Date Received
- October 31, 2023
- Date of Event
- October 20, 2023
- Report Date
- July 19, 2024
- Manufacturer
- PHILIPS MEDICAL SYSTEMS
- Product Code
- DSI
- UDI-DI
- 00884838082236
- PMA / PMN Number
- K113125
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
- Health Professional
- Yes
Narratives
PHILIPS IS IN PROCESS OF OBTAINING ADDITIONAL INFORMATION. A FINAL REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION.
D4: PRODUCT UDI - THE MANUFACTURING DATE FOR THE REPORTED DEVICE IS PRIOR TO 24SEPT2016, THEREFORE NO UDI.
DIAGNOSTIC/FUNCTIONAL TESTING WAS PERFORMED AT THE PHILIPS AUTHORIZED REPAIR FACILITY. RESULTS OF FUNCTIONAL TESTING INDICATE THAT THE SPEAKER PRODUCED SOUND. BASED ON THE INFORMATION AVAILABLE AND THE TESTING CONDUCTED WE WERE UNABLE TO REPLICATE THE REPORTED PROBLEM. THE REPORTED PROBLEM WAS NOT CONFIRMED. ALTHOUGH THE SPEAKER WAS CONFIRMED TO BE FUNCTIONING PER SPECIFICATION DURING TESTING IT WAS INDICATED THAT THERE WAS NO SOUND AT THE TIME OF THE EVENT, THE SPEAKER HAS BEEN REPLACED PER CURRENT PROCESS. THE DEVICE WAS OPERATIONAL AFTER REPAIRS WERE COMPLETED. THE INVESTIGATION CONCLUDES THAT NO FURTHER ACTION IS REQUIRED AT THIS TIME. IF ADDITIONAL INFORMATION IS RECEIVED THE COMPLAINT FILE WILL BE REOPENED.
THE CUSTOMER BIOMED REPORTED THAT THE FRONT BEZEL IS BROKEN, THERE ARE NO SOUNDS COMING FROM THE DEVICE DURING BOOTUP, AND THE TOUCHSCREEN IS INTERMITTENT. THE DEVICE IS PENDING BENCH REPAIR. THE DEVICE WAS NOT IN USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2125016 | 4003409 | 4003409 | DSI | PHILIPS MEDICAL SYSTEMS | 865350 | 00884838082236 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |