FDA Adverse Event Malfunction Summary report: N

I-STAT TROPONIN (CTNI) CARTRIDGE

MDR report key: 18038119 · Received October 31, 2023

Report

Report Number
2245578-2023-00148
Event Type
Malfunction
Date Received
October 31, 2023
Date of Event
October 6, 2023
Report Date
November 14, 2023
Manufacturer
ABBOTT POINT OF CARE INC.
Product Code
MMI
UDI-DI
10054749000187
PMA / PMN Number
K031739
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

APOC INCIDENT # (B)(4). APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.

Additional Manufacturer Narrative · 0

APOC INCIDENT: # (B)(4). THE INVESTIGATION WAS COMPLETED ON 08-NOV-2023. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) CONFIRMED THE CARTRIDGE LOT MET FINISHED GOODS (FG) RELEASE CRITERIA. RETAINED CARTRIDGE TESTING MET THE ACCEPTANCE CRITERIA OUTLINED IN APPENDIX 1 OF Q04.01.003 REV. AL (PRODUCT COMPLAINT LEVEL 2 AND LEVEL 3 INVESTIGATION PROCEDURE). NO DEFICIENCY HAS BEEN IDENTIFIED FOR CTNI CARTRIDGE LOT B23124.

Description of Event or Problem · 0

ON 17-OCT-2023, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER REGARDING I-STAT TROPONIN CARTRIDGES THAT YIELDED A FALSE POSITIVE DISCREPANT RESULT ON A PATIENT. THERE WAS NO ADDITIONAL PATIENT INFORMATION AT THE TIME OF THIS REPORT. RETURN PRODUCT IS NOT AVAILABLE FOR INVESTIGATION. METHOD : RESULT : SAMPLE: I-STAT , NO RESULT PROVIDED , NI. I-STAT , <0.02 NG/ML , B. NO TEST/COLECT TIMES PROVIDED. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON THE LIMITED INFORMATION AVAILABLE THAT SUGGESTED THAT PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY. THERE IS NO EVIDENCE THE PATEINT SUFFERED AN MI. THE INVESTIGATION IS UNDERWAY. PER I-STAT SYSTEM MANUAL: ART: 715595-00V. REPORTABLE RANGE: 0.00 - 50.00. REFERENCE RANGE: 0.00 - 0.03 (REPRESENTS THE 0 TO 97.5% RANGE OF RESULTS). REFERENCE RANGE: 0.00 - 0.08 (REPRESENTS THE 0 TO 99% RANGE OF RESULTS).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2124832 I-STAT TROPONIN (CTNI) CARTRIDGE CTNI CARTRIDGE MMI ABBOTT POINT OF CARE INC. NA B23124 10054749000187

Patients

Seq Age Sex Outcome Treatment
1 71 YR Female