MONOPLUS VIOLET 1 (4) 70CM HR26S (M)RCP
Report
- Report Number
- 3003639970-2023-00350
- Event Type
- Malfunction
- Date Received
- October 31, 2023
- Report Date
- October 31, 2023
- Manufacturer
- B. BRAUN SURGICAL, S.A.
- Product Code
- NEW
- PMA / PMN Number
- SEE H10
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
G4: REPORTED DEVICE NOT MARKETED IN THE U.S., HOWEVER, SIMILAR DEVICES OR DEVICES THAT SHARE COMPONENTS, RAW MATERIALS, PROCESS METHODS OR OTHER TECHNOLOGICAL CHARACTERISTICS ARE REGISTERED WITHIN THE U.S. RELATED 510K NUMBER K031216. ANALYSIS AND RESULTS: THERE ARE NO PREVIOUS COMPLAINTS OF THIS CODE-BATCH. WE MANUFACTURED AND DISTRIBUTED IN THE MARKET (B)(4) UNITS OF THIS CODE-BATCH. THERE ARE NO UNITS IN OUR STOCK. WE HAVE RECEIVED ONE OPEN SAMPLE TO ANALYZE THIS COMPLAINT, THE NEEDLE IS MISSING AND THE THREAD IS NOT WOUND ON THE PACK. HOWEVER, WITHOUT ANY CLOSED SAMPLE WE CANNOT CARRY OUT AN ANALYSIS IN ORDER TO TAKE A DECISION. REVIEWED THE BATCH MANUFACTURING RECORD, THIS BATCH HAD AN INCIDENCE BUT WAS RELEASED TO THE MARKET FULFILLING USP/EP AND B.BRAUN SURGICAL REQUIREMENTS. FINAL CONCLUSION: IN SPITE OF RECEIVING A DEFECTIVE SAMPLE, WITHOUT CLOSED SAMPLES A SUITABLE ANALYSIS CANNOT BE PERFORMED. NEVERTHELESS, WE TAKE NOTE OF THIS INCIDENCE AND IF ANY CLOSED SAMPLE IS RECEIVED IN THE FUTURE, WE WILL RE-OPEN THE CASE AND ANALYSE IT. WE REGRET ANY INCONVENIENCE THIS ISSUE MAY HAVE CAUSED AND THANK YOU FOR YOUR COLLABORATION. ACTIONS ON PRODUCT DISTRIBUTED OF THIS REFERENCE/BATCH: BASED ON THE CONCLUSION DERIVED FROM INVESTIGATION, IT IS NOT REQUIRED TO MAKE ACTIONS IN DISTRIBUTED PRODUCT. CORRECTIVE/PREVENTIVE ACTIONS: ACCORDING TO OUR INTERNAL PROCEDURES, THERE IS NO NEED TO ESTABLISH CORRECTIVE OR PREVENTIVE ACTIONS. NEVERTHELESS, THIS COMPLAINT IS RECORDED FOR TRENDING ANALYSIS TO ASSESS IF ACTIONS ARE NEEDED IN THE FUTURE.
IT WAS REPORTED AN ISSUE WITH MONOPLUS SUTURE. THE CLIENT REPORTED THAT THE THREAD COMES OFF NEEDLE TOO EASILY. NO FURTHER INFORMATION HAS BEEN RECEIVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1942358 | MONOPLUS VIOLET 1 (4) 70CM HR26S (M)RCP | SYNTHETIC ABSORBABLE MONOFILAR SUTURES | NEW | B. BRAUN SURGICAL, S.A. | C2024088 | 122261 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |