FDA Adverse Event Malfunction Summary report: N

RAD-97 PULSE CO-OXIMETER

MDR report key: 18036242 · Received October 30, 2023

Report

Report Number
3019388613-2023-00265
Event Type
Malfunction
Date Received
October 30, 2023
Date of Event
September 28, 2023
Report Date
October 6, 2023
Manufacturer
MASIMO - 15750 ALTON PKWY
Product Code
MWI
UDI-DI
00843997012317
PMA / PMN Number
K212161
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
BIOMEDICAL ENGINEER
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL MANUFACURING NARRATIVE: OTHER, OTHER TEXT: THE RETURNED DEVICE WAS EVALUATED. THE DEVICE POWERED ON WITH A BLANK SCREEN AND ISSUED AN AUDIBLE AND VISUAL ALARM, HOWEVER THE DEVICE DID NOT LOSE POWER DURING TESTING. AN IMPROPERLY SEATED FLEX CABLE AND LOOSE TAPE AT THE AT CONNECTIVITY BOARD CAUSED THE BLANK DISPLAY AND ALARMS.

Additional Manufacturer Narrative · 0

ADDITIONAL MANUFACURING NARRATIVE: ATTEMPTS HAVE BEEN MADE TO OBTAIN THE PRODUCT. THE PRODUCT HAS NOT BEEN RETURNED TO MASIMO TO ALLOW AN ANALYSIS TO BE PERFORMED. IF THE PRODUCT IS RETURNED FOR EVALUATION OR NEW INFORMATION IS OBTAINED, A FOLLOW UP REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

THIS CORRECTION UPDATES THE UDI INFORMATION, 510K NUMBER, AND MODEL NUMBER TO ENSURE ALIGNMENT WITH GUDID. THE 510K NUMBER IN GUDID WAS UPDATED TO K212161 TO CORRESPOND WITH MASIMO'S 4-DIGIT MODEL NUMBER (9738).

Description of Event or Problem · 0

THE CUSTOMER REPORTED THAT THE RAD-97 UNIT WILL POWER ON, BUT WILL LOSE POWER QUICKLY. THERE WERE NO PATIENT IMPACT OR CONSEQUENCES REPORTED.

Description of Event or Problem · 0

THE CUSTOMER REPORTED THAT THE RAD-97 UNIT WILL POWER ON, BUT WILL LOSE POWER QUICKLY. THERE WERE NO PATIENT IMPACT OR CONSEQUENCES REPORTED.

Description of Event or Problem · 0

THE CUSTOMER REPORTED THAT THE RAD-97 UNIT WILL POWER ON, BUT WILL LOSE POWER QUICKLY. THERE WERE NO PATIENT IMPACT OR CONSEQUENCES REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1949798 RAD-97 PULSE CO-OXIMETER MONITOR, PHYSIOLOGICAL, PATIENT (WITHOUT ARRHYTHMIA DETECTION OR ALARMS) MWI MASIMO - 15750 ALTON PKWY 9738 00843997012317

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown