FDA Adverse Event Malfunction Summary report: N

CS100 INTRA-AORTIC BALLOON PUMP, FRENCH, 220V

MDR report key: 18034432 · Received October 30, 2023

Report

Report Number
2249723-2023-04644
Event Type
Malfunction
Date Received
October 30, 2023
Date of Event
October 20, 2023
Report Date
July 1, 2025
Manufacturer
DATASCOPE CORP.
Product Code
DSP
UDI-DI
10607567107509
PMA / PMN Number
K063525
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
BIOMEDICAL ENGINEER
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

UPDATED DATA: B4, E1(NAME),(IMPACT). CORRECTED DATA: E1 ( EMAIL).

Additional Manufacturer Narrative · 0

A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF OUR INVESTIGATION.

Additional Manufacturer Narrative · 0

UDI RELATED DATA QUALITY UPDATES ONLY: PROVIDING UPDATED DEVICE IDENTIFICATION INFORMATION IN ALIGNMENT WITH GUDID. SUSPECT DEVICE ORIGINALLY DISTRIBUTED AS A SYS98XT, UPGRADED TO CS300 USING CONVERSION KIT. NO UDI IS PROVIDED AS PRODUCT WAS DISCONTINUED PRIOR TO GUDID IMPLEMENTATION TIMELINE.

Additional Manufacturer Narrative · 0

CORRECTED FIELD: D1, D4 (# UDI), G4 (510 K), H6 (MEDICAL DEVICE ¿ PROBLEM CODE). UPDATED FIELDS: B4, D8, D9, G3, G6, H2, H3, H6 (INVESTIGATION FINDINGS, TYPE OF INVESTIGATION, INVESTIGATION CONCLUSIONS, COMPONENT CODE), H11. A GETINGE FIELD SERVICE ENGINEER (FSE) STATES, SCREEN PROBLEM NOT OBSERVED, DONE DISASSEMBLY OF THE SCREEN CASINGS. DISASSEMBLY OF THE SCREEN CABLE PRESENT. INSTALLATION OF THE NEW CABLE (D012-00-1059). REASSEMBLY OF COVER OK. FUNCTIONAL TEST OK. ELECTRICAL TEST OK. EVERYTHING WORKS CORRECTLY.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING START UP, THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) UNIT HAD A SCREEN PROBLEM. THERE WAS NO PATIENT INVOLVEMENT.

Description of Event or Problem · 0

N/A.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1942067 CS100 INTRA-AORTIC BALLOON PUMP, FRENCH, 220V SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL DSP DATASCOPE CORP. 0998-00-0446-64 10607567107509

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown