CS100 INTRA-AORTIC BALLOON PUMP, FRENCH, 220V
Report
- Report Number
- 2249723-2023-04644
- Event Type
- Malfunction
- Date Received
- October 30, 2023
- Date of Event
- October 20, 2023
- Report Date
- July 1, 2025
- Manufacturer
- DATASCOPE CORP.
- Product Code
- DSP
- UDI-DI
- 10607567107509
- PMA / PMN Number
- K063525
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- BIOMEDICAL ENGINEER
- Health Professional
- Yes
Narratives
UPDATED DATA: B4, E1(NAME),(IMPACT). CORRECTED DATA: E1 ( EMAIL).
A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF OUR INVESTIGATION.
UDI RELATED DATA QUALITY UPDATES ONLY: PROVIDING UPDATED DEVICE IDENTIFICATION INFORMATION IN ALIGNMENT WITH GUDID. SUSPECT DEVICE ORIGINALLY DISTRIBUTED AS A SYS98XT, UPGRADED TO CS300 USING CONVERSION KIT. NO UDI IS PROVIDED AS PRODUCT WAS DISCONTINUED PRIOR TO GUDID IMPLEMENTATION TIMELINE.
CORRECTED FIELD: D1, D4 (# UDI), G4 (510 K), H6 (MEDICAL DEVICE ¿ PROBLEM CODE). UPDATED FIELDS: B4, D8, D9, G3, G6, H2, H3, H6 (INVESTIGATION FINDINGS, TYPE OF INVESTIGATION, INVESTIGATION CONCLUSIONS, COMPONENT CODE), H11. A GETINGE FIELD SERVICE ENGINEER (FSE) STATES, SCREEN PROBLEM NOT OBSERVED, DONE DISASSEMBLY OF THE SCREEN CASINGS. DISASSEMBLY OF THE SCREEN CABLE PRESENT. INSTALLATION OF THE NEW CABLE (D012-00-1059). REASSEMBLY OF COVER OK. FUNCTIONAL TEST OK. ELECTRICAL TEST OK. EVERYTHING WORKS CORRECTLY.
IT WAS REPORTED THAT DURING START UP, THE CS300 INTRA-AORTIC BALLOON PUMP (IABP) UNIT HAD A SCREEN PROBLEM. THERE WAS NO PATIENT INVOLVEMENT.
N/A.