FDA Adverse Event Injury Summary report: N

ATTAIN OTW

MDR report key: 1803288 · Received August 17, 2010

Report

Report Number
2649622-2010-08821
Event Type
Injury
Date Received
August 17, 2010
Date of Event
April 30, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P010015/S003
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TU
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS LIMITED DUE TO CONFIDENTIALITY CONCERNS. THIS MODEL NUMBER IS NOT APPROVED FOR DISTRIBUTION IN THE UNITED STATES, HOWEVER, IT IS SIMILAR TO A DEVICE MARKETED IN THE U.S. THE EVENT IS BEING REPORTED DUE TO AN ALLEGED OR CONFIRMED MALFUNCTION. CORRECTION: FURTHER INFORMATION INDICATED THAT THE THREE LEADS ARE STILL IN USE SO THE EVENT DESCRIPTION UPDATED, EXPLANT DATES REMOVED AND DEVICE CODES CHANGED. EVALUATION SUMMARY: (B)(4) ACTUAL LONGEVITY IS < 80% OF 99.9% LONGEVITY LIMIT. THE DEVICE WAS FULLY FUNCTIONAL WITH NO HIGH CURRENT DRAIN OR EVIDENCE OF BATTERY PROBLEMS. WITHOUT THE HISTORY OF THE PROGRAMMED SETTINGS THROUGHOUT ITS SERVICE LIFE, THERE IS NO WAY TO DETERMINE WHY THE LONGEVITY DID NOT MATCH THE PREDICTED MODEL.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS LIMITED DUE TO CONFIDENTIALITY CONCERNS. THIS MODEL NUMBER IS NOT APPROVED FOR DISTRIBUTION IN THE UNITED STATES, HOWEVER, IT IS SIMILAR TO A DEVICE MARKETED IN THE U.S. THE EVENT IS BEING REPORTED DUE TO AN ALLEGED OR CONFIRMED MALFUNCTION. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE HAD REACHED ERI (ELECTIVE REPLACEMENT INDICATOR). THE PHYSICIAN THOUGHT THAT THE DEVICE REACHED ERI BEFORE THE PREDICTED TIME FRAME. THE DEVICE WAS EXPLANTED AND REPLACED. IT WAS FURTHER REPORTED THAT THE LEADS WERE STILL IN USE WITH THE NEW DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE HAD REACHED ERI (ELECTIVE REPLACEMENT INDICATOR). THE PHYSICIAN THOUGHT THAT THE DEVICE REACHED ERI BEFORE THE PREDICTED TIME FRAME. THE DEVICE AND THE LEADS WERE EXPLANTED AND REPLACED. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ATTAIN OTW IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 4193 ASKU

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention