FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 1803274 · Received August 17, 2010

Report

Report Number
2649622-2010-08817
Event Type
Injury
Date Received
August 17, 2010
Date of Event
April 15, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Removal / Correction Number
ASKU
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) NO ANOMALIES FOUND; FULL LEAD WAS RETURNED AND ANALYZED. OUTER TUBING OVERLAY WAS BREACHED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WAS A RV LEAD INTEGRITY WARNING. ADDITIONAL INFORMATION OBTAINED THROUGH FOLLOW UP INDICATE THERE WAS OVERSENSING AND SHORT INTERVAL COUNTS THAT LED TO REPOSITIONING PROCEDURE ON THE DAY AFTER THE IMPLANT. IT WAS FURTHER REPORTED THAT THE PHYSICIAN DISCOVERED A NICK IN THE INSULATION IN THE PROXIMAL PORTION OF THE LEAD POSSIBLY INDICATING DAMAGE AT THE IMPLANT TIME. THE LEAD WAS WAS REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WAS A RV LEAD INTEGRITY WARNING. THE LEAD WAS REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947 ASKU

Patients

Seq Age Sex Outcome Treatment
1 45 YR Required Intervention D224DRG IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 5076 IMPLANTABLE PACING LEAD