PRECISION XTRA/OPTIUM
Report
- Report Number
- 2954323-2007-16270
- Event Type
- Malfunction
- Date Received
- August 31, 2007
- Date of Event
- August 2, 2007
- Report Date
- August 31, 2007
- Manufacturer
- ABBOTT DIABETES CARE LIMITED UK
- Product Code
- NBW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- UNKNOWN
Narratives
(B)(4). THE DEVICE WAS RETURNED AND THE INVESTIGATION REVEALED THE COMPLAINT WAS NOT CONFIRMED AND NO NEW ISSUES WERE OBSERVED. ALL RESULTS WERE WITHIN THE RANGE SPECIFICATION AND NO ERRORS ARE OBSERVED DURING CONTROL SOLUTION TESTING. THE READINGS REPORTED BY THE CUSTOMER WERE NOT FOUND IN THE METER'S INTERNAL MEMORY LOG.
CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR PRECISION XTRA/OPTIUM BLOOD GLUCOSE METER. CUSTOMER REPORTED RECEIVING READINGS OF 52 MG/DL AND 400 MG/DL WITHIN 10 MINUTES. ALL TESTS WERE PERFORMED ON THE FINGER. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PRECISION XTRA/OPTIUM | BLOOD GLUCOSE METER | NBW | ABBOTT DIABETES CARE LIMITED UK | NI | 42205 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |