FDA Adverse Event Malfunction Summary report: N

ATTAIN ABILITY

MDR report key: 1802956 · Received August 17, 2010

Report

Report Number
2649622-2010-08757
Event Type
Malfunction
Date Received
August 17, 2010
Date of Event
June 17, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
OJX
PMA / PMN Number
P080006
Removal / Correction Number
ASKU
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED AND ANALYZED. PRIMARY FINDING: NO ANOMALIES FOUND. ALL CONDUCTORS HAD BLOOD/BODY FLUID (NOT OBSTRUCTED) AND THE LEAD WAS STRETCHED. VISUAL ANALYSIS OF THE RETURNED LEAD SUGGESTS DAMAGE AT IMPLANT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN IMPLANT ATTEMPT, THE LEFT VENTRICULAR LEAD HAD HIGH THRESHOLDS. ANOTHER LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ATTAIN ABILITY IMPLANTABLE PACING LEAD OJX MEDTRONIC PUERTO RICO, INC. 4196 ASKU

Patients

Seq Age Sex Outcome Treatment
1 50 YR Required Intervention