HAMILTON-G5
Report
- Report Number
- 3001421318-2023-03683
- Event Type
- Malfunction
- Date Received
- October 30, 2023
- Date of Event
- October 25, 2023
- Report Date
- January 23, 2025
- Manufacturer
- HAMILTON MEDICAL AG
- Product Code
- CBK
- UDI-DI
- 07630002800013
- PMA / PMN Number
- K193228
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- 505
Narratives
ALARM LAMP BOARD WAS FOUND DEFECTIVE AND REPLACED.
ALARM LAMP BOARD WAS FOUND DEFECTIVE AND REPLACED. HAMILTON MEDICAL AG COMPLAINT NUMBER: (B)(4). FOLLOW-UP 1 - CORRECTED INFORMATION: **UDI RELATED DATA QUALITY UPDATES ONLY** FIELD D4 WAS UPDATED WITH FULL UDI INFORMATION AS REQUESTED. UPDATED FIELDS: B4, D1, D4, G3, G6, H2, H4.
ALARM LAMP BOARD WAS FOUND DEFECTIVE AND REPLACED. ADDITIONAL INFORMATION ADDED IN FOLLOW-UP #2 CER 143499. THE ISSUE WAS DISCOVERED DURING SERVICE MAINTENANCE WITH NO PATIENT INVOLVED. CONSEQUENTLY, THE EVENT DID NOT CAUSE OR CONTRIBUTED TO A DEATH OR SERIOUS INJURY. THE MOST PROBABLE ROOT CAUSE OF THE LED FAILURES WAS CHEMICAL CORROSION DUE TO THE GAS PERMEABILITY. THE PARTNER SERVICE TECHNICIAN REPLACED THE ALARM LAMP BOARD PN: 159272 TO SOLVE THE ISSUE AND PERFORMED THE COMPLETE SERVICE SOFTWARE. THE VENTILATOR WAS THEN RELEASED FOR USE ON THE PATIENT. THE ISSUE IS CONSIDERED A REPORTABLE EVENT BECAUSE IT COULD POTENTIALLY IMPACT THE PATIENT SAFETY DURING VENTILATION SINCE MEDICAL PERSONNEL MAY NOT BE VISUALLY ALERTED AS INTENDED/NEEDED (HOWEVER AUDIBLE ALARMS ARE WORKING).
HAMILTON MEDICAL AG RECEIVED THE FOLLOWING INCIDENT DESCRIPTION: RED AND AMBER LEDS ARE NOT LIGHTING UP.
HAMILTON MEDICAL AG RECEIVED THE FOLLOWING INCIDENT DESCRIPTION: RED AND AMBER LEDS ARE NOT LIGHTING UP.
HAMILTON MEDICAL AG RECEIVED THE FOLLOWING INCIDENT DESCRIPTION: RED AND AMBER LEDS ARE NOT LIGHTING UP.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2009368 | HAMILTON-G5 | HAMILTON-G5 VENTILATOR | CBK | HAMILTON MEDICAL AG | 159002 | 07630002800013 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |