FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 1802773 · Received August 17, 2010

Report

Report Number
2649622-2010-08687
Event Type
Malfunction
Date Received
August 17, 2010
Date of Event
June 18, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Removal / Correction Number
ASKU
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) HELIX DISENGAGED FROM HELICAL CHANNEL. THERE WAS BLOOD/BODY FLUID IN/ON THE HELIX MECHANISM (SLEEVE HEAD) AND DISTAL CONDUCTOR (NOT OBSTRUCTED). FULL LEAD RETURNED AND ANALYZED.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ANALYSIS OF THE LEAD IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING IMPLANT ATTEMPT, THE HELIX MECHANISM STOPPED WORKING. THERE WAS POSITIONING/FIXATION DIFFICULTY. THE LEAD WAS REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947 ASKU

Patients

Seq Age Sex Outcome Treatment
1 77 YR Required Intervention