FDA Adverse Event Injury Summary report: N

CONCERTO CRT-D DR

MDR report key: 1802772 · Received August 17, 2010

Report

Report Number
6000144-2010-04189
Event Type
Injury
Date Received
August 17, 2010
Date of Event
June 30, 2010
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
NIK
PMA / PMN Number
P010031/S031
Removal / Correction Number
Z-0126-2010
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL. THE SPRINT FIDELIS LEAD MODEL IS INCLUDED IN A FIELD ADVISORY. EVALUATION SUMMARY: (B)(4) A HIGH CURRENT DRAIN CONDITION WAS FOUND DURING DEVICE ANALYSIS. ELECTRICAL ANALYSIS FOUND THE CAUSE OF THE HIGH CURRENT DRAIN TO BE CURRENT LEAKAGE IN THE BATTERY FILTER CAPACITORS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WAS EARLY BATTERY DEPLETION. THE DEVICE WAS EXPLANTED AND REPLACED. IT WAS ALSO REPORTED THAT THE LEFT VENTRICULAR LEAD WAS PUT IN THE RV PORT OF THE DEVICE AND THE PACE/SENSE PORTION OF THE HIGH VOLTAGE RIGHT VENTRICULAR LEAD WAS IN THE LV PORT OF THE DEVICE. THE LEADS REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONCERTO CRT-D DR IMPLANTABLE PACEMAKER/CARDIO/DEFIB NIK MEDTRONIC MED REL, INC. C154DWK ASKU

Patients

Seq Age Sex Outcome Treatment
1 83 YR Required Intervention 5076 IMPLANTABLE PACING LEAD