FDA Adverse Event Injury Summary report: N

ATTAIN OTW

MDR report key: 1802766 · Received August 17, 2010

Report

Report Number
2649622-2010-08691
Event Type
Injury
Date Received
August 17, 2010
Date of Event
June 21, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P010015/S003
Removal / Correction Number
ASKU
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THE LV LEAD THRESHOLD WAS 6V. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ATTAIN OTW IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 4193 ASKU

Patients

Seq Age Sex Outcome Treatment
1 67 YR Other C154DWK IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 6947 IMPLANTABLE TACHY LEAD| 5076 IMPLANTABLE PACING LEAD