- Report Number
- 2649622-2010-08734
- Event Type
- Malfunction
- Date Received
- August 17, 2010
- Date of Event
- June 30, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S039
- Removal / Correction Number
- ASKU
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) HELIX DISENGAGED FROM HELICAL CHANNEL. DISTAL CONDUCTOR HAD BLOOD/BODY FLUID (NOT OBSTRUCTED). THERE WAS BLOOD IN/ON THE HELIX MECHANISM (SLEEVE HEAD). FULL LEAD RETURNED AND ANALYZED.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ANALYSIS OF THE LEAD IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.
IT WAS REPORTED THAT AFTER ONE LEAD REPOSITIONING, THE HELIX DID RETRACT BUT WOULD NOT EXTEND A SECOND TIME AFTER MULTIPLE TURNS. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
IT WAS REPORTED THAT AFTER 1 LEAD REPOSITIONING, THE HELIX DID RETRACT BUT WOULD NOT EXTEND A SECOND TIME AFTER MULTIPLE TURNS. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT QUATTRO SECURE S | IMPLANTABLE TACHY LEAD | LWS | MEDTRONIC PUERTO RICO, INC. | 6935 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 53 YR | Required Intervention |