CAPSUREFIX NOVUS
Report
- Report Number
- 2649622-2010-08735
- Event Type
- Injury
- Date Received
- August 17, 2010
- Date of Event
- June 14, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P930039/S017
- Removal / Correction Number
- ASKU
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CORRECTION: CHANGE EVENT LOCATION FROM UNK TO HOSPITAL. EVALUATION SUMMARY: (B)(4) NO ANOMALIES FOUND. OUTER INSULATION BREACHED CUT AND BLOOD IN/ON HELIX MECHANISM. FULL LEAD RETURNED AND ANALYZED.
IT WAS REPORTED THAT THE LEAD WAS REPOSITIONED SEVERAL TIMES AND HAD HIGH THRESHOLDS IN ALL LOCATIONS. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSUREFIX NOVUS | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC PUERTO RICO, INC. | 4076 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 79 YR | Required Intervention | 4076 IMPLANTABLE PACING LEAD| VEDR01 IMPLANTABLE PULSE GENERATOR |