FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX NOVUS

MDR report key: 1802681 · Received August 17, 2010

Report

Report Number
2649622-2010-08737
Event Type
Malfunction
Date Received
August 17, 2010
Date of Event
June 22, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S009
Removal / Correction Number
ASKU
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) A BENT TIP ELECTRODE WAS OBSERVED; DAMAGED AT IMPLANT. THE FULL LEAD WAS RETURNED AND ANALYZED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SCREW MECHANISM WAS NOT WORKING PROPERLY ON THE LEAD. THE DOCTOR EXERCISED THE LEAD BEFORE IMPLANTING, AND THE LEAD WAS NEVER IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5076 ASKU

Patients

Seq Age Sex Outcome Treatment
1 82 YR Other