FDA Adverse Event Injury Summary report: N

ACHIEVA 1.5T NOVA

MDR report key: 1802424 · Received August 9, 2010

Report

Report Number
3003768277-2010-00151
Event Type
Injury
Date Received
August 9, 2010
Date of Event
July 11, 2010
Report Date
July 29, 2010
Manufacturer
PHILIPS MEDICAL SYSTEMS
Product Code
LNH
PMA / PMN Number
K052013
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INVESTIGATION IS STILL ONGOING ON THIS EVENT. WHEN THE INVESTIGATION IS COMPLETED A FOLLOW-UP WILL BE SENT TO THE FDA.

Description of Event or Problem · 1

THE PATIENT HAD A MR EXAM. SHE WAS SCANNED IN THE HEAD FIRST SUPINE POSITION WITH THE FOUR CHANNEL SHOULDER COIL. HER RIGHT ARM WAS AT HER SIDE. WITHIN 48 HOURS AFTER THE EXAMINATION A LARGE 2ND TO 3RD DEGREE BURN WAS FOUND ON HER INNER FOREARM NEAR THE ELBOW. THE INTERFACE BOX WAS DIRECTLY IN CONTACT WITH THE PT'S ARM DURING THE EXAMINATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ACHIEVA 1.5T NOVA LNH (MAGNETIC RESONANCE DIAGNOSTIC DEVICE) LNH PHILIPS MEDICAL SYSTEMS 781196 NA

Patients

Seq Age Sex Outcome Treatment
1 67 YR Other