FDA Adverse Event Malfunction Summary report: N

DEPTH GAUGE FOR 2.0MM AND 2.4MM SCREWS

MDR report key: 18020524 · Received October 27, 2023

Report

Report Number
8030965-2023-13488
Event Type
Malfunction
Date Received
October 27, 2023
Date of Event
October 6, 2023
Report Date
October 27, 2023
Manufacturer
SYNTHES GMBH
Product Code
HTJ
UDI-DI
10886982189943
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. H3, H4, H6: PART # 319.006, SYNTHES LOT # H873103, SUPPLIER LOT # H873103, RELEASE TO WAREHOUSE DATE:27 APR 2020, SUPPLIER: (B)(4). NO NCR'S GENERATED DURING PRODUCTION. REVIEW OF THE DEVICE HISTORY RECORD(S) SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THIS PRODUCT, AND ANY SUB-COMPONENTS, WHICH WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. THE PRODUCT WAS RETURNED TO DEPUY SYNTHES FOR EVALUATION. THE DEPUY SYNTHES TEAM CONDUCTED A VISUAL INSPECTION OF THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED SAMPLE REVEALED THAT THE NEEDLE OF DEPTH GAUGE FOR 2.0MM AND 2.4MM SCREWS BROKE OFF. THE BROKEN FRAGMENT WAS RETURNED FOR EVALUATION. BASED ON THE OBSERVED CONDITION OF THE DEVICE THE INVESTIGATION WAS ABLE TO CONFIRM THE REPORTED EVENT. A DIMENSIONAL INSPECTION WAS NOT PERFORMED AS IT IS NOT APPLICABLE TO THE COMPLAINT CONDITION. THE OBSERVED CONDITION OF THE DEVICE WAS CONSISTENT WITH A RANDOM COMPONENT FAILURE THAT MAY HAVE BEEN CAUSED BY EXPOSURE TO UNINTENDED FORCES. AS PART OF DEPUY SYNTHES QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. THE OVERALL COMPLAINT WAS CONFIRMED AS THE OBSERVED CONDITION OF THE DEPTH GAUGE FOR 2.0MM AND 2.4MM SCREWS WOULD CONTRIBUTE TO THE COMPLAINED DEVICE ISSUE. THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE COMPLAINT CONDITION. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. D9: COMPLAINANT PART IS EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION, BUT HAS YET TO BE RECEIVED. H3, H4, H6: THE PRODUCT WAS NOT RETURNED TO DEPUY SYNTHES, HOWEVER PHOTOS WERE PROVIDED FOR REVIEW. THE PHOTO INVESTIGATION REVEALED THAT NEEDLE OF THE DEPTH GAUGE FOR 2.0MM AND 2.4MM SCREWS WAS BROKEN. THE OBSERVED CONDITION OF THE DEVICE WAS CONSISTENT WITH A RANDOM COMPONENT FAILURE THAT MAY HAVE BEEN CAUSED BY EXPOSURE TO UNINTENDED FORCES. ¿ ¿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

Description of Event or Problem · 0

IT WAS REPORTED THAT ON OCTOBER 6, 2023, THE DEPTH GAUGE WAS BROKEN - THE TIP WAS STUCK. IT WAS DISCOVERED DURING EQUIPMENT CLEANING. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT INVOLVES ONE DEPTH GAUGE FOR 2.0MM AND 2.4MM SCREWS. THIS IS REPORT 1 OF 1 FOR (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1023546 DEPTH GAUGE FOR 2.0MM AND 2.4MM SCREWS DEPTH GAUGE HTJ SYNTHES GMBH H873103 10886982189943

Patients

Seq Age Sex Outcome Treatment
1 Unknown