ACCESS® 2 IMMUNOASSAY SYSTEM

MDR report key: 1802047 · Received August 16, 2010

Report

Report Number
2122870-2010-00432
Event Type
Malfunction
Date Received
August 16, 2010
Date of Event
July 19, 2010
Report Date
August 16, 2010
Manufacturer
BECKMAN COULTER INC.
Product Code
MMI
PMA / PMN Number
K922823/A007
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE SAMPLE WAS COLLECTED IN A PST TUBE. QC WAS WITHIN THE CUSTOMER'S ESTABLISHED RANGES PRIOR TO THE EVENT. A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED: THE FSE PERFORMED A DIAGNOSTIC TESTING WHICH MET SPECIFICATIONS. NO HARDWARE ISSUES WERE NOTED THAT WOULD HAVE CONTRIBUTED TO THIS EVENT. A DEFINITIVE ROOT CAUSE HAS NOT BEEN DETERMINED FOR THIS EVENT.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC., (BCI) REGARDING TROPONIN (ACCU TNI) RESULTS IN THE RISK STRATIFICATION RANGE AND ABOVE THE AMI CUT-OFF GENERATED BY THE ACCESS 2 IMMUNOASSAY SYSTEM FOR ONE PATIENT. SUBSEQUENT TESTING ON THIS INSTRUMENT AND ON AN ALTERNATE METHOD RESULTED IN THE NORMAL REFERENCE RANGE. THE RESULTS WERE REPORTED OUT OF THE LAB. NO REPORTS OF DEATH, INJURY, OR CHANGE TO PATIENT TREATMENT HAVE BEEN REPORTED IN CONNECTION WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ACCESS® 2 IMMUNOASSAY SYSTEM DISCRETE PHOTOMETRIC CHEMISTRY ANALYZER MMI BECKMAN COULTER INC. ACCESS 2