FDA Adverse Event Malfunction Summary report: N

AUTOMATED PD SET W/CASSETTE4 PRONG

MDR report key: 1802039 · Received August 16, 2010

Report

Report Number
1423500-2010-02563
Event Type
Malfunction
Date Received
August 16, 2010
Date of Event
July 26, 2010
Report Date
July 27, 2010
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). A REVIEW OF ALL BATCH RECORD DOCUMENTS WAS PERFORMED WITH NO ISSUES NOTED DURING THE MANUFACTURING PROCESS. THERE WERE NO DEVIATIONS FROM STANDARD PROCEDURE. THE SAMPLE WAS DISCARDED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW UP MDR WILL BE SENT.

Additional Manufacturer Narrative · 1

(B)(4). AN ENGINEERING QUALITY REVIEW WAS COMPLETED FOR THIS REPORT OF DAMAGED TUBING. THE REPORT WAS NOT CONFIRMED DUE TO A LACK OF SAMPLE AND THE ROOT CAUSE WAS UNDETERMINED. A BATCH REVIEW WAS PERFORMED ON THE ASSOCIATED LOT WITH NO ISSUES NOTED. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.

Description of Event or Problem · 1

IT IS REPORTED THAT THE PATIENT WAS REVISED FOR LOOSENING AND PAIN.

Description of Event or Problem · 1

THIS REPORT ADDRESSES SAMPLE PRODUCT 1 OF 15 REPORTED (B)(6) 2010. A HOME CARE SERVICE REPRESENTATIVE CONTACTED GLOBAL FIELD SURVEILLANCE TO REPORT A HOME PATIENT (HP) RECEIVED A CASSETTE WHERE THE TUBING WAS CRUSHED AND UNUSABLE. ON (B)(6) 2010 PRODUCT SURVEILLANCE SPOKE WITH THE HP WHO STATED THE DAMAGE WAS FOUND BEFORE USING THE PRODUCT. ON (B)(6) 2010 PRODUCT SURVEILLANCE SPOKE WITH THE HP WHO STATED THE EXTERNAL BOX OF THIS PRODUCT LOT WAS NOT DAMAGED. THE HP STATED APPROXIMATELY HALF OF THE CASE OF THE CASSETTES (15) HAD TO BE DISCARDED DUE TO CRUSHED TUBING. THE HP STATED IN THIS PARTICULAR LOT, THE 15 CASSETTES HE DISPOSED OF ALL HAD 'U' SHAPE DAMAGE AND THE TUBING WAS UNUSABLE. THE HP STATED IT WAS A VARIETY OF LINES DAMAGED; THERE WAS NO TREND OF SPECIFIC LINE. HOWEVER, THE HP STATED ALL LINES WERE CRUSHED ABOVE THE CLAMPS. THE HP STATED IT APPEARED THE CLAMPS CAUSED THE DAMAGE TO THE TUBING BY HOW THE CASSETTES WERE PACKAGED. THE HP STATED HE HAS HAD NO ISSUE WITH HIS NEW LOT OF SUPPLIES AND HAS SELDOM HAD ISSUES WITH PRODUCT IN THE PAST. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION. NO FURTHER INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AUTOMATED PD SET W/CASSETTE4 PRONG SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME H10E29048

Patients

Seq Age Sex Outcome Treatment
1 70 YR