FDA Adverse Event Malfunction Summary report: N

INPEN MMT-105NNBLNA NOVO NORDISK BLUE

MDR report key: 18020253 · Received October 27, 2023

Report

Report Number
3012822846-2023-01776
Event Type
Malfunction
Date Received
October 27, 2023
Date of Event
July 1, 2023
Report Date
December 7, 2023
Manufacturer
COMPANION MEDICAL INC
Product Code
FMF
UDI-DI
000010862088000344
PMA / PMN Number
K201337
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE INPEN PAIRED TO THE COMMERCIAL APP. INPEN PASSED BASELINE AND WIRELESS FUNCTIONALITY. APP LOGBOOK DISPLAYED: 15U, 15U, 15U, 15U, 15U, 15U, 15U, 15U, 15U, 15U. PERFORMED LEADSCREW RESET TORQUE TEST. INPEN PASSED AND IS WITHIN SPECIFICATION (CW: 0.50 OZF-IN AND CCW: 4.40 OZF-IN). INPEN PASSED FRONT CAP INVESTIGATION. PENDING FURTHER INVESTIGATION PERFORMED IN SAN DIEGO LOCATION. IN CONCLUSION: PER SAN DIEGO ANALYSIS: INPEN PASSED BASE LINE FUNCTIONALITY TEST AND DISPLACEMENT DOSE ACCURACY. INPEN PASSED LEADSCREW RESET TORQUE (CW 2.77 AND CCW 2.38). PAIRED TO COMMERCIAL APP USING 11 UNITS. INPEN BASELINE AND WIRELESS FUNCTIONALITY. APP LOGBOOK DISPLAYED: 15,15,15,15,15,15,15,15,15,15. KNOB WAS CUT TO REVEAL ELECTRONICS HOUSING. NO SIGNS OF ABRASION WERE FOUND. ENCODER BASE BOND WAS INTACT. NO PROBLEMS FOUND WITH THIS INPEN ALL FUNCTIONS TESTED OK. DIFFICULT TO DIAL/DOSE WAS NOT CONFIRMED. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Additional Manufacturer Narrative · 0

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Description of Event or Problem · 0

INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE CUSTOMER REPORTED THAT THE INPEN BUTTON WAS WIGGLY EVERY TIME. TROUBLESHOOTING WAS PERFORMED AND FOUND THAT THE CUSTOMER REPORTED DOSE KNOB/DIAL WAS DIFFICULT TO TURN. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE CUSTOMER WILL DISCONTINUE THE USE OF THE INPEN AND WILL BE RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1700208 INPEN MMT-105NNBLNA NOVO NORDISK BLUE SYRINGE, PISTON FMF COMPANION MEDICAL INC MMT-105NNBLNA B0599 000010862088000344

Patients

Seq Age Sex Outcome Treatment
1 12 YR Male