FDA Adverse Event Malfunction Summary report: N

COLLEAGUE CX VOLUMETRIC INFUSION PUMP SINGLE CHANNEL

MDR report key: 1802017 · Received August 16, 2010

Report

Report Number
6000001-2010-02605
Event Type
Malfunction
Date Received
August 16, 2010
Date of Event
July 14, 2010
Report Date
July 15, 2010
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K063696
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).EVALUATION SUMMARY:THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WHICH OVER-DELIVERED WAS NOT CONFIRMED NOR DUPLICATED DURING PRODUCT EVALUATION. DELIVERY ACCURACY TESTS WERE PERFORMED AND FOUND THE PUMP TO BE DELIVERING WITHIN SPECIFICATIONS. THEREFORE, NO ASSIGNABLE CAUSE COULD BE PROVIDED FOR THE REPORTED CONDITION AND NO REPAIR WAS NECESSARY.SHOULD ADDITIONAL INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.

Description of Event or Problem · 1

THE FACILITY REPORTED A COLLEAGUE INFUSION PUMP WHICH OVER-INFUSED DURING BIO-MEDICAL TESTING. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.THIS DEVICE IS A REMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 6.13.90.