OT PING METER
Report
- Report Number
- 2939301-2010-06662
- Event Type
- Malfunction
- Date Received
- August 16, 2010
- Report Date
- July 26, 2010
- Manufacturer
- LIFESCAN, INC.
- Product Code
- MDS
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
THE LAY USER/PATIENT'S PRODUCT(S) HAS BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS WITH THE FOLLOWING FINDINGS: THE METER INVOLVED IN THIS CASE FAILED TESTING. SPC TEST PORT FOUND DIRTY. IF ANY ADDITIONAL INFORMATION IS AVAILABLE, THE FDA WILL BE NOTIFIED IN A SECOND FOLLOW UP REPORT. AT THIS TIME, WE CONSIDER THIS MATTER CLOSED.
LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT.
A PATIENT REPORTED BLOOD GLUCOSE RESULTS OF "73 MG/DL" WITH A LIFESCAN METER AND "42 MG/DL" ON A LABORATORY DEVICE, PERFORMED WITHIN 10 MINUTES OF EACH OTHER. BETWEEN THE TESTS THERE WAS NO INTERVENTION THAT WOULD BE EXPECTED TO CHANGE THE BLOOD GLUCOSE. BASED ON STATISTICAL METHODOLOGY, THE CALCULATED DIFFERENCE OF THESE GLUCOSE RESULTS EXCEEDS LIFESCAN'S CRITERIA FOR ACCURACY, GREATER THAN 1 MG/DL (0.056 MMOL/L) PER MINUTE AND GREATER THAN 20 MG/DL (1.11 MMOL/L) OR 20%. THE ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WERE NO ALLEGATIONS OF HARM OR INJURY. REPLACEMENT PRODUCTS WERE SENT TO THE PATIENT.
THE LAY USER/PATIENT CONTACTED LIFESCAN ALLEGING THE METER DISPLAYS ERROR 5 MESSAGE. THE ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WERE NO ALLEGATIONS OF HARM OR INJURY. REPLACEMENT PRODUCTS WERE SENT TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OT PING METER | GLUCOSE MONITORING SYS/KIT | MDS | LIFESCAN, INC. | 3027684 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 62 YR |