FDA Adverse Event Malfunction Summary report: N

MINIMED 720G OUS BLE CONNECT 3.0 MGDL

MDR report key: 18017871 · Received October 27, 2023

Report

Report Number
2032227-2023-291932
Event Type
Malfunction
Date Received
October 27, 2023
Date of Event
October 3, 2023
Report Date
December 21, 2023
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
UDI-DI
000000763000541347
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PL
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

UNIT PASSED THE SELF TEST, DISPLACEMENT TEST, REWIND TEST, PRIME/SEATING TEST, BASIC OCCLUSION TEST, FORCE SENSOR TEST AND OCCLUSION TEST. TEST P-CAP AND RESERVOIR LOCKED PROPERLY INTO RESERVOIR COMPARTMENT DURING TESTING. NO UNEXPECTED INSULIN FLOW BLOCKED ALARMS NOTED DURING TESTING. SUCCESSFULLY UTILIZED THUMP TO DOWNLOAD HISTORY/TRACE FILES. THE FOLLOWING ALARMS/ALERTS WERE NOTED ON EVENT DATE: 12 NO DELIVERY ALARMS STARTING ON 10/01/2023 02:31:00.000 ENDING ON 10/01/2023 02:31:00.000. THE FOLLOWING NO DELIVERY ALARMS WERE NOTED ON EVENT DATE: 1 NO DELIVERY ALARM DURING PRIMING: START DATE 10/01/2023 02:31:00.000 2 NO DELIVERY ALARM DURING BOLUS: START DATE 10/01/2023 02:39:14.000 END DATE 10/01/2023 02:41:02.000 2 NO DELIVERY ALARM DURING PRIMING: START DATE 10/01/2023 05:01:00.000 END DATE 10/01/2023 08:48:00.000 1 NO DELIVERY ALARM DURING BOLUS: START DATE 10/01/2023 09:52:47.000 1 NO DELIVERY ALARM DURING BASAL: START DATE 10/01/2023 13:16:00.000 4 NO DELIVERY ALARM DURING PRIMING: START DATE 10/01/2023 13:17:00.000 END DATE 10/01/2023 19:43:00.000 1 NO DELIVERY ALARM DURING BOLUS: START DATE 10/02/2023 14:46:53.000. PUMP WAS CUT OPEN TO PERFORM VISUAL INSPECTION AND FOUND NO EVIDENCE OF PHYSICAL OR MOISTURE DAMAGE¿ON THE ELECTRONIC ASSEMBLY, MOTOR, OR FORCE SENSOR. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: PILLOWING KEYPAD OVERLAY. PUMP PASSED FUNCTION TESTING. NO UNEXPECTED INSULIN FLOW BLOCKED ALARMS NOTED DURING TESTING. NO DELIVERY ALARM NOT CONFIRMED. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Additional Manufacturer Narrative · 0

THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED NORTHRIDGE SITE, PER VARIANCE 5. CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. SELECT PATIENT INFORMATION CANNOT BE PROVIDED DUE TO REGIONAL PRIVACY REGULATIONS. THE REPORTED DEVICE IS NOT MARKETED IN THE UNITED STATES, BUT IT IS A SAME/SIMILAR DEVICE TO ONE THAT IS MARKETED OUTSIDE THE UNITED STATES. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Description of Event or Problem · 0

INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE CUSTOMER REPORTED AN INSULIN FLOW BLOCK. DURING BOLUS TROUBLESHOOTING WAS PERFORMED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE CUSTOMER WILL DISCONTINUE USING THE INSULIN PUMP. THE PRODUCT WILL BE RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1150204 MINIMED 720G OUS BLE CONNECT 3.0 MGDL PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-1810 NG3357798H 000000763000541347

Patients

Seq Age Sex Outcome Treatment
1 38 YR Unknown