FDA Adverse Event Malfunction Summary report: N

CELL-DYN SAPPHIRE VENT HEAD ASSEMBLY

MDR report key: 1801362 · Received August 16, 2010

Report

Report Number
2919069-2010-00381
Event Type
Malfunction
Date Received
August 16, 2010
Report Date
July 25, 2010
Manufacturer
ABBOTT DIAGNOSTICS DIVISION
Product Code
GKZ
PMA / PMN Number
D019916
Removal / Correction Number
2919069-7/26/10-005-R
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). CONCOMITANT MEDICAL DEVICE: CELL-DYN SAPPHIRE ANALYZER, LIST # 8H00-01, (B)(4). THE CAUSE OF THE CELL-DYN SAPPHIRE VENT HEAD ASSEMBLY ASPIRATION ERROR (UNDETECTED SHORT SAMPLE) WAS DUE TO A DEFECTIVE CELL-DYN VENT NEEDLE FROM THE SUPPLIER. A PRODUCT RECALL INFORMED ABBOTT CUSTOMERS TO DISCARD ANY CELL-DYN SAPPHIRE VENT HEAD ASSEMBLIES AND REPLACE WITH A NEW CELL-DYN VENT HEAD ASSEMBLY PROVIDED TO THE CUSTOMER WITH THE RECALL LETTER.

Description of Event or Problem · 1

THE CUSTOMER OBSERVED SLUGGISH MOVEMENT OF THE CELL-DYN SAPPHIRE ANALYZER'S VENT HEAD ASSEMBLY. ABBOTT FIELD SERVICE REPRESENTATIVE SERVICED THE ANALYZER AND FOUND THE VENT HEAD ASSEMBLY DEFECTIVE, THEREFORE, IT WAS REPLACED. THE ANALYZER'S CONTROLS AND SAMPLES WERE ALL OK FOLLOWING SERVICE. THERE WAS NO IMPACT TO PATIENT MANAGEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CELL-DYN SAPPHIRE VENT HEAD ASSEMBLY VENT NEEDLE USED ON THE CELL-DYN ANALYZER GKZ ABBOTT DIAGNOSTICS DIVISION

Patients

Seq Age Sex Outcome Treatment
1 CELL-DYN SAPPHIRE ANALYZER, LIST # 8H00-01