FDA Adverse Event Injury Summary report: N

TALENT THORACIC STENT GRAFT SYSTEM

MDR report key: 1801282 · Received August 11, 2010

Report

Report Number
2953200-2010-01530
Event Type
Injury
Date Received
August 11, 2010
Date of Event
July 12, 2010
Report Date
July 12, 2010
Manufacturer
MEDTRONIC CARDIOVASCULAR
Product Code
MIH
PMA / PMN Number
P070007
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4): EVALUATION RESULTS AND CONCLUSIONS: THE PT'S BLOOD PRESSURE WAS NOT LOWERED, PRIOR TO TREATMENT.

Description of Event or Problem · 1

A TALENT STENT GRAFT SYSTEM WAS IMPLANTED IN A PT FOR THE ENDOVASCULAR TREATMENT OF A 6 CM THORACIC ANEURYSM. PRIOR TO THE STENT GRAFT IMPLANT THERE WAS A COIL EMBOLIZATION PERFORMED AT THE LSA, AND BYPASS SURGERY BETWEEN CAROTID TO CAROTID. IT WAS REPORTED THAT A TALENT 3434 STENT GRAFT WAS IMPLANTED PROXIMALLY AT THE OSTIUM OF INNOMINATE ARTERY, HOWEVER, DUE TO PT'S BLOOD PRESSURE THE STENT GRAFT WAS PUSHED DOWN TO THE DISTAL END OF THE THORACIC ANEURYSM (MFR# 2953200-2010-01528). THE PT'S BLOOD PRESSURE WAS LOWERED BY ATP, AND THE PHYSICIAN PLACED A TALENT 3636 STENT GRAFT PROXIMALLY TO THE TF3434. DURING EXPANDING AND POSITIONING OF TF3636 STENT GRAFT, AND BEFORE COMPLETE DEPLOYMENT, THERE WAS A RETURN OF SPONTANEOUS CIRCULATION, SO TF3636 MOVED DISTALLY AND THE ANEURYSM WAS NOT SEALED (MFR# 2953200-2010-01529). ANOTHER PHYSICIAN ASSISTED AND ONCE AGAIN THE PT'S BLOOD PRESSURE WAS LOWERED BY ATP, AND AN ATTEMPT TO PLACE A TALENT 3838 STENT GRAFT PROXIMALLY TF3636. DURING THE REMOVAL OF THE TF3838, THE GAP BETWEEN THE TAPERED TIP AND THE EDGE OF OUTER SHEATH OF DELIVERY CATHETER WAS CAUGHT ON THE APEX STENT OF THE EXISTING IMPLANTED TALENT 3636 STENT GRAFT. THE PHYSICIAN EXCHANGED THE GUIDEWIRE, HOWEVER, STILL WAS UNABLE TO ADVANCE THE DEVICE TO THE INTENDED LANDING ZONE. THE PHYSICIAN ELECTED TO PULL THE STENT GRAFT DISTALLY AND PLACE THE TF3838 STENT GRAFT AT AN UNINTENDED LOCATION. THE PHYSICIAN INSERTED ANOTHER MFR'S STENT GRAFT ADVANCING IT THROUGH THE PREVIOUSLY PLACED TALENT 3636 AND TALENT 3434. THE OTHER MFR'S STENT GRAFT WAS PLACED PROXIMAL TO THE TF3636 DEVICE, HOWEVER, THERE WAS A PROXIMAL TYPE I ENDOLEAK. THE PHYSICIAN PLACED A TALENT 3434 STENT GRAFT PROXIMALLY TO AND THE ENDOLEAK WAS RESOLVED. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PT IS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TALENT THORACIC STENT GRAFT SYSTEM MIH MEDTRONIC CARDIOVASCULAR NA UNK

Patients

Seq Age Sex Outcome Treatment
1 79 YR Required Intervention