FDA Adverse Event Malfunction Summary report: N

CARTO® 3 SYSTEM

MDR report key: 18010571 · Received October 26, 2023

Report

Report Number
2029046-2023-02433
Event Type
Malfunction
Date Received
October 26, 2023
Date of Event
August 21, 2023
Report Date
October 26, 2023
Manufacturer
BIOSENSE WEBSTER INC
Product Code
DQK
UDI-DI
10846835000870
PMA / PMN Number
K213264
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

CORRECTION TO SECTION G4. PMA/ 510(K) K213264. FULL UDI NUMBER IS NOT CURRENTLY AVAILABLE AS WE ARE PENDING THE MANUFACTURE DATE OF THE DEVICE. IF ADDITIONAL INFORMATION IS RECEIVED REGARDING THIS EVENT, A SUPPLEMENTAL 3500A REPORT WILL BE SUBMITTED TO THE FDA. MANUFACTURER'S REF. NO: (B)(4).

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION WAS RECEIVED, THE H4. DEVICE MANUFACTURE DATE 25-OCT-2024. THEREFORE, THE SECTION D4. PRIMARY UDI NUMBER (B)(4) WAS UPDATED. IF ADDITIONAL INFORMATION IS RECEIVED REGARDING THIS EVENT, A SUPPLEMENTAL 3500A REPORT WILL BE SUBMITTED TO THE FDA. MANUFACTURER'S REF. NO: (B)(4).

Additional Manufacturer Narrative · 0

DEVICE EVALUATION DETAILS: A BIOSENSE WEBSTER INC (BWI) FIELD SERVICE ENGINEER (FSE) SPOKE WITH THE BWI REPRESENTATIVE WHO CONFIRMED THE ORDER OF EVENTS AND THE BURNING SMELL. THE FSE WENT ONSITE TO REPLACE THE PATIENT INTERFACE UNIT (PIU) AND LOCATION PAD (LP). FSE PERFORMED ACCEPTANCE TESTING PROCEDURE (ATP) AFTER REPLACEMENT. SYSTEM PASSED TESTING AND IS READY FOR USE. THE REPLACED PIU/LP KIT WAS SENT TO THE BWI'S REPAIR CENTER AND WAS SERVICED. PIU MAIN MODULE AND DC/DC CARD WERE REPLACED BECAUSE OF ELECTRICAL BURN DAMAGES. PER TECHNICAL SERVICES (TS), BOTH MAIN MODULE AND DC/DC CARD SHOWED SIGNS OF BURN. THESE CARDS HAD BEEN SCRAPPED. ALSO, THE LOCATION PAD WAS REPLACED DUE TO PHYSICAL DAMAGE. THE SYSTEM PASSED ALL TESTS AFTER REPAIR AND PUT BACK IN STOCK. PHOTOS OF THE REPLACED PIU MAIN MODULE AND DC/DC CARD WERE PROVIDED TO THE DEVICE MANUFACTURE. NO SIGNS OF BURN WERE CONFIRMED AFTER REVIEWING OF THE PROVIDED PICTURES. NO ADDITIONAL INVESTIGATION CAN BE PERFORMED AS THE REPLACED PARTS ARE NOT AVAILABLE ANYMORE. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE CARTO 3 SYSTEM # 11444, AND NO INTERNAL ACTIONS RELATED TO THE REPORTED COMPLAINT CONDITION WERE IDENTIFIED. THE HISTORY OF CUSTOMER COMPLAINTS REPORTED DURING THE LAST YEAR AND ASSOCIATED WITH CARTO 3 SYSTEM # 11444 WAS REVIEWED. NO SIMILAR ADDITIONAL COMPLAINT WAS FOUND. IF ADDITIONAL INFORMATION IS RECEIVED REGARDING THIS EVENT, A SUPPLEMENTAL 3500A REPORT WILL BE SUBMITTED TO THE FDA. MANUFACTURER'S REF. # (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT A PATIENT UNDERWENT A ATRIAL FLUTTER RIGHT (R-AFL) ABLATION PROCEDURE WITH A CARTO®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

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2111241 CARTO® 3 SYSTEM COMPUTER, DIAGNOSTIC, PROGRAMMABLE DQK BIOSENSE WEBSTER INC 10846835000870

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown