FDA Adverse Event Injury Summary report: N

COMPRESSION ANASTOMOSIS RING CAR 27/COLONRING

MDR report key: 1800942 · Received August 10, 2010

Report

Report Number
3005278776-2010-00120
Event Type
Injury
Date Received
August 10, 2010
Date of Event
June 30, 2010
Report Date
July 14, 2010
Manufacturer
NITI SURGICAL SOLUTIONS LTD.
Product Code
FZP
PMA / PMN Number
K062008
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS SUBMITTED ON BEHALF OF THE MANUFACTURER (NITI SURGICAL SOLUTIONS LTD; 3005278776) AND THE (B)(4), AS NITI SURGICAL SOLUTIONS LTD SERVES AS THE DESIGNATED COMPLAINT HANDLING UNIT FOR BOTH FACILITIES. THE PRODUCTION HISTORY FILE WAS REVIEWED, INDICATING THAT THE DEVICE WAS RELEASED ACCORDING TO THE PRODUCT SPECIFICATIONS. THE RING WAS RETURNED AND EVALUATED. NO FAILURE WAS DETECTED AND THE RING WAS FOUND TO BE WITHIN ITS SPECIFICATION. ANASTOMOTIC LEAKAGE IS ONE OF THE MOST COMMON COMPLICATIONS OF COLORECTAL ANASTOMOTIC PROCEDURES AND THE CURRENT CUMULATIVE LEAK RATE FOUND WITH THE USE OF THE CAR DEVICE IS WITHIN THE RANGE REPORTED IN THE LITERATURE FOR ANASTOMOSIS DEVICES.

Description of Event or Problem · 1

A PATIENT SUFFERING FROM RECTAL CANCER AND CROHN'S DISEASE UNDERWENT LEFT HEMICOLECTOMY PROCEDURE. SIDE TO END ANASTOMOSIS WAS CREATED WITH THE CAR DEVICE. THE ANASTOMOSIS BROKE DOWN AFTER 5 DAYS. HARTMANN WITH PROTECTIVE COLOSTOMY WAS PERFORMED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COMPRESSION ANASTOMOSIS RING CAR 27/COLONRING IMPLANTABLE CLIP (FZP) FZP NITI SURGICAL SOLUTIONS LTD. CAR 27/COLONRING 43271134T

Patients

Seq Age Sex Outcome Treatment
1 55 YR Required Intervention