REPLY
Report
- Report Number
- 1000165971-2010-00857
- Event Type
- Injury
- Date Received
- August 10, 2010
- Date of Event
- July 16, 2010
- Report Date
- July 21, 2010
- Manufacturer
- SORIN BIOMEDICA C.R.M., S.R.L.
- Product Code
- NVZ
- PMA / PMN Number
- P950029
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). THIS EVENT CONCERNS A DEVICE THAT WAS MANUFACTURED AND USED OUTSIDE THE UNITED STATES. ANALYSIS IS PENDING.
REPORTEDLY, A HIGH VENTRICULAR IMPEDANCE (OVER 3000 OHMS) WAS MEASURED DURING THE FOLLOW-UP ONE WEEK AFTER IMPLANTATION; IN ADDITION, PACING FAILURE WAS OBSERVED AT THE HIGHEST OUTPUT FROM THE PACEMAKER, AND THE R WAVE AMPLITUDE WENT DOWN FROM 7-8MV (AS MEASURED AT IMPLANT) TO 2-3MV. AN X-RAY WAS TAKEN, WHICH REVEALED THAT THE TIP OF THE LEAD CONNECTOR PIN WAS COMING OUT SLIGHTLY FROM THE PACEMAKER CONNECTOR PORT. THEREFORE A REINTERVENTION WAS CARRIED OUT: THE LEAD WAS CONFIRMED TO BE FIXED WITHIN THE PACEMAKER PORT BY A PULL TEST; THEREFORE, THE SETSCREW WAS UNSCREWED IN ORDER TO TAKE THE LEAD OUT FROM THE PORT. A PSA WAS UTILIZED TO OBTAIN THE LEAD MEASUREMENTS, WHICH REVEALED NO ABNORMAL RESULT. A NEW PACEMAKER WAS SUCCESSFULLY IMPLANTED. THE PACEMAKER INVOLVED IN THIS MDR REPORT WAS RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | REPLY | NVZ | SORIN BIOMEDICA C.R.M., S.R.L. | REPLY DR | 2435 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |