FDA Adverse Event Injury Summary report: N

SYNCHROMED II

MDR report key: 1800892 · Received August 10, 2010

Report

Report Number
3004209178-2010-06009
Event Type
Injury
Date Received
August 10, 2010
Date of Event
July 1, 2010
Report Date
July 13, 2010
Manufacturer
MDT PUERTO RICO OPERATIONS CO., JUNCOS
Product Code
LKK
PMA / PMN Number
P860004
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

(B)(4)

Description of Event or Problem · 1

ON (B)(6)-2010, IT WAS REPORTED THAT A PATIENT EXPERIENCED INCREASED BASELINE SPASTICITY AND A RETURN OF THE FOLLOWING SYMPTOMS: IRRITABILITY, RASH, AND ITCHING. THE PATIENT RESPONDED TO A 30 MCG BOLUS, BUT THE SYMPTOMS REAPPEARED. A SECOND 30 MCG BOLUS WAS GIVEN AND THE PATIENT RESPONDED AGAIN. A DYE STUDY WAS PERFORMED BUT THE CATHETER COULD NOT BE ASPIRATED. A SECOND DYE STUDY WAS PERFORMED THE NEXT DAY, WHICH ALSO RESULTED IN AN INABILITY TO ASPIRATE FROM THE CATHETER. THIS TIME THE PHYSICIAN PUSHED THROUGH ANYWAY, AND THE DYE HAD FLOWED THROUGH NICELY AND APPEARED IN THE INTRATHECAL SPACE. A CT SCAN ALSO SHOWED GOOD FLOW OF THE DYE INTRATHECALLY. ON (B)(6)-2010, THE PATIENT RETURNED TO THE HOSPITAL WITH INCREASED SPASTICITY, CLONUS, AND SPASMS. WITHDRAWAL WAS ALSO REPORTED. THE CATHETER WAS NOTED AS FRACTURED, AND A REPLACEMENT/REVISION SURGERY WAS SCHEDULED. ON (B)(6)-2010, A PUMP ISSUE OCCURRED, AND MOTOR STALL WAS DISCUSSED. A ROLLER STUDY WAS PERFORMED AND REPEATED, WHICH SUGGESTED THAT THERE WAS NO MOVEMENT OF THE PUMP ROLLERS. A SINGLE BOLUS OF 30 MCG OVER 10 MINUTES WAS PERFORMED, AND ROLLER MOVEMENT WAS DETECTED. THE PHYSICIAN HAD DECIDED TO REPLACE THE PUMP ALONG WITH THE CATHETER. THE PATIENT WAS NOTED AS DOING WELL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SYNCHROMED II LKK MDT PUERTO RICO OPERATIONS CO., JUNCOS 8637-40 NA

Patients

Seq Age Sex Outcome Treatment
1 15 YR Required Intervention IMPLANTED:| CATHETER: MODEL 8709SC, LOT# N192954005| EXPLANTED: