FDA Adverse Event Malfunction Summary report: N

ETHILON BLK 10CM M0.2

MDR report key: 18007725 · Received October 25, 2023

Report

Report Number
2210968-2023-08178
Event Type
Malfunction
Date Received
October 25, 2023
Date of Event
September 29, 2023
Report Date
December 14, 2023
Manufacturer
ETHICON INC.
Product Code
GAR
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # (B)(4). DATE SENT TO THE FDA: 12/14/2023. H6 COMPONENT CODE: G07002. NO DEVICE PROBLEM FOUND. H3 INVESTIGATIONAL SUMMARY: THE PRODUCT WAS RETURNED TO ETHICON FOR EVALUATION. VISUAL INSPECTION AND FUNCTIONAL TESTING WERE CONDUCTED ON THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED SAMPLE REVEALED THAT SEVEN PACKETS WERE RETURNED FOR ANALYSIS. UPON INITIAL INSPECTION, OF THE SAMPLES, NO EXTERNAL DAMAGES WERE OBSERVED ON THE PACKETS. IN ORDER TO EVALUATE THE CONDITIONS OF THE RETURNED SAMPLES, THE PACKETS WERE OPENED, AND NO DEFECTS WERE DETECTED. THE SWAGE AND ATTACHMENT AREA WERE NOTED TO BE AS EXPECTED. THE SUTURES WERE EXAMINED ALONG OF THE STRAND AND NO ISSUES RELATED TO BREAKAGE SUTURES OR ANOMALIES WERE OBSERVED DURING THE EVALUATION. THE FUNCTIONAL TEST WAS PERFORMED USING INSTRON EQUIPMENT, AND THE PULL FORCE AND TENSILE STRENGTH WERE ABOVE THE MINIMUM REQUIREMENTS. THE EVENT DESCRIBED COULD NOT BE CONFIRMED AS THE DEVICE PERFORMED WITHOUT ANY DEFECT NOTED. IT SHOULD BE NOTED THAT AS PART OF OUR QUALITY PROCESS, EACH BATCH IS RANDOMLY INSPECTED AND FUNCTIONALLY TESTED DURING MANUFACTURING TO ENSURE THE DEVICE MEETS THE REQUIRED SPECIFICATIONS PRIOR TO SHIPMENT. ADDITIONAL COMPLAINT INFORMATION MONITORING FOR POTENTIAL SAFETY SIGNALS IS CONDUCTED THROUGH COMPLAINT TRENDING AS PART OF POST-MARKET SURVEILLANCE. THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY ETHICON INC, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, ETHICON INC OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. RELATED REPORTS: 2210968-2023-08177, 2210968-2023-08179, 2210968-2023-08180, 2210968-2023-08181.

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # ==> (B)(4) H6 COMPONENT CODE: G07002 - DEVICE NOT RETURNED THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY ETHICON INC, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, ETHICON INC OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. ATTEMPTS HAVE BEEN MADE TO RETRIEVE THE DEVICE. TO DATE THE DEVICE HAS NOT BEEN RETURNED. IF THE DEVICE OR FURTHER DETAILS ARE RECEIVED AT A LATER DATE A SUPPLEMENTAL MEDWATCH WILL BE SENT. ATTEMPTS ARE BEING MADE TO OBTAIN THE FOLLOWING INFORMATION. TO DATE NO RESPONSE HAS BEEN PROVIDED. IF FURTHER DETAILS ARE RECEIVED AT A LATER DATE A SUPPLEMENTAL MEDWATCH WILL BE SENT. - PLEASE PERFORM AND DOCUMENT THE FOLLOW UP ATTEMPT FOR PRODUCT RETURN. PLEASE DOCUMENT THE SHIPMENT TRACKING NUMBER IN NOTES OR RMAO SECTIONS. - PLEASE PROVIDE THE SOURCE OR NAME OF PERSON PROVIDING ANSWERS TO FOLLOW-UP QUESTIONS (NOT THE PERSON RELAYING/SUBMITTING ANSWERS TO LOC OR CHU). A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE LOT, AND NO NON-CONFORMANCES WERE IDENTIFIED. EVENTS REPORTED VIA: (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT A PATIENT UNDERWENT A SCLERA PROCEDURE ON (B)(6) 2023 AND SUTURE WAS USED. DURING THE PROCEDURE, THE CUSTOMER REPORTED SUTURE BREAKAGE INTRO-OPERATIVE. A NEW PACKAGE WITH A DIFFERENT LOT AND FINISHED THE PROCEDURE. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1700716 ETHILON BLK 10CM M0.2 SUTURE, NONABSORBABLE, SYNTHETIC GAR ETHICON INC. TCMRDQ

Patients

Seq Age Sex Outcome Treatment
1 Unknown