FDA Adverse Event Malfunction Summary report: N

PRECISION XTRA/OPTIUM

MDR report key: 1800607 · Received September 7, 2007

Report

Report Number
2954323-2007-16680
Event Type
Malfunction
Date Received
September 7, 2007
Date of Event
August 9, 2007
Report Date
September 7, 2007
Manufacturer
ABBOTT DIABETES CARE LIMITED UK
Product Code
NBW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE ((B)(4)) WAS RETURNED AND THE COMPLAINT WAS NOT CONFIRMED AND NO NEW ISSUES WERE OBSERVED DURING THE PRODUCT TESTING. ALL RESULTS WERE WITHIN THE RANGE SPECIFICATION AND NO ERRORS WERE OBSERVED DURING CONTROL SOLUTION TESTING. THE READINGS REPORTED WERE PRESENT IN THE METER'S MEMORY LOG.

Description of Event or Problem · 1

CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR PRECISION XTRA BLOOD GLUCOSE METER. CUSTOMER REPORTED RECEIVING READINGS OF 27 MG/DL, 44 MG/DL AND 88 MG/DL WITHIN 10 MINUTES. ALL TESTS WERE PERFORMED ON THE FINGER. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PRECISION XTRA/OPTIUM BLOOD GLUCOSE METER NBW ABBOTT DIABETES CARE LIMITED UK NI 42062

Patients

Seq Age Sex Outcome Treatment
1 UNK