FDA Adverse Event Malfunction Summary report: N

AQUABPLUS, HF 3000

MDR report key: 18006061 · Received October 25, 2023

Report

Report Number
3010850471-2023-00082
Event Type
Malfunction
Date Received
October 25, 2023
Date of Event
July 21, 2022
Report Date
July 31, 2024
Manufacturer
VIVONIC GMBH
Product Code
FIP
PMA / PMN Number
K133829
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
EC
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

PLANT INVESTIGATION: PHYSICAL EVALUATION OF A SAMPLE WAS NOT REQUIRED FOR THIS CASE. THE REPORTED ISSUE WAS CONFIRMED USING THE PROVIDED INFORMATION. THE CAUSE OF THE REPORTED FAILURE WAS DETERMINED TO BE A BAD ELECTRICAL CONTACT OF THE WIRING AT THE MOTOR PROTECTION SWITCH. THE DAMAGE RESULTED FROM EITHER (1) THE ELECTRICAL CONTACT AT THE MOTOR PROTECTION SWITCH PUMP P1(S) RUNNING WITH ONLY TWO LINE CONDUCTORS OR (2) THE INNER BIMETAL CONTACT OF THE MOTOR PROTECTION TRIPPING FROM INCREASED TEMPERATURE CAUSED BY THE TRANSITION RESISTANCE OF THE ELECTRICAL CONTACT. BOTH ISSUES RESULT IN HIGHER CURRENTS AND HIGHER THERMAL ENERGY, AND ULTIMATELY THERMAL DECOMPOSITION. THIS FAILURE IS A KNOWN ISSUE. A ROOT CAUSE ANALYSIS IS IN PROGRESS AND CORRECTIVE/PREVENTIVE ACTIONS WILL BECOME AVAILABLE WITH AN IMPROVED DESIGN FOR THE ELECTRICAL MONITOR, WHICH INCLUDES THE WIRING AT THE MAIN SWITCH (THE MOTOR PROTECTION SWITCH IN THE CURRENT DESIGN). THE IMPROVED DESIGN IS CURRENTLY NOT AVAILABLE FOR THE AFFECTED MARKET SEGMENT, BUT THE RELEASE IS PLANNED BY COMPLETION OF THE DESIGN CHANGE. ACCORDING TO THE PROVIDED INFORMATION, THE MOTOR PROTECTION SWITCH, CONTACTOR AND THE WIRING WERE REPLACED TO SOLVE THE ISSUE. THE CURRENT STATUS OF THE MACHINE IS UNKNOWN. IT IS RECOMMENDED, IF NOT ALREADY DONE, TO REPLACE THE WIRING AND THE MOTOR PROTECTION SWITCH IN STAGE 1 AND STAGE 2 WITH THE AVAILABLE DESIGN. ONLY INSTALLATION OF MANUFACTURER SPARE PARTS IS ADVISED.

Additional Manufacturer Narrative · 0

THE PLANT INVESTIGATION IS IN PROCESS. A SUPPLEMENTAL MDR WILL BE SUBMITTED UPON COMPLETION OF THIS ACTIVITY.

Description of Event or Problem · 0

IT WAS REPORTED TO FRESENIUS THAT THERMAL DAMAGE WAS IDENTIFIED WITHIN THE AQUABPLUS REVERSE OSMOSIS (RO) SYSTEM. THE DAMAGE WAS RELATED TO THE IGNITION SWITCH GROUP, ELECTRICAL CONTACTOR OF THE PRESSURE PUMP, AND THE ELECTRICAL WIRING OF THE CONNECTION BETWEEN THE MAIN IGNITION SWITCH AND THE ELECTRICAL CONTACTOR THAT FEEDS THE PRESSURE PUMP OF THE RO SYSTEM. FAILURE DUE TO THE EFFECT OF HEAT DAMAGE WAS EVIDENT FROM A VISUAL INSPECTION. THERE WAS "EVIDENCE OF OVERHEATING IN THE CABLES" AS WELL AS SULFATED TERMINALS. THERE WAS NO PATIENT INVOLVEMENT NOR REPORTED PERSONAL HARM TO ANYONE AS A RESULT OF ENCOUNTERING THE REPORTED THERMAL DAMAGE. WIRING REPLACEMENT OF THE PROTECTION SWITCH AND MAIN PUMP CONTACTOR, WHICH INCLUDED THE SULFATED TERMINALS AND FAULTY CONTACTORS, WAS COMPLETED TO RESOLVE THE REPORTED THERMAL DAMAGE. THE CABLE GAUGE WAS ALSO REPLACED IN ORDER TO SUPPORT A GREATER LOAD OF CURRENT INTENSITY. IT WAS REPORTED THAT THE MACHINE HAS APPROXIMATELY 5,320 HOURS. IT WAS NOT REPORTED THAT ANY SAMPLES ARE AVAILABLE TO BE RETURNED TO THE MANUFACTURER FOR PHYSICAL EVALUATION. THE MACHINE DATA IS AVAILABLE FOR REVIEW BY THE MANUFACTURER.

Description of Event or Problem · 0

IT WAS REPORTED TO FRESENIUS THAT THERMAL DAMAGE WAS IDENTIFIED WITHIN THE AQUABPLUS REVERSE OSMOSIS (RO) SYSTEM. THE DAMAGE WAS RELATED TO THE IGNITION SWITCH GROUP, ELECTRICAL CONTACTOR OF THE PRESSURE PUMP, AND THE ELECTRICAL WIRING OF THE CONNECTION BETWEEN THE MAIN IGNITION SWITCH AND THE ELECTRICAL CONTACTOR THAT FEEDS THE PRESSURE PUMP OF THE RO SYSTEM. FAILURE DUE TO THE EFFECT OF HEAT DAMAGE WAS EVIDENT FROM A VISUAL INSPECTION. THERE WAS "EVIDENCE OF OVERHEATING IN THE CABLES" AS WELL AS SULFATED TERMINALS. THERE WAS NO PATIENT INVOLVEMENT NOR REPORTED PERSONAL HARM TO ANYONE AS A RESULT OF ENCOUNTERING THE REPORTED THERMAL DAMAGE. WIRING REPLACEMENT OF THE PROTECTION SWITCH AND MAIN PUMP CONTACTOR, WHICH INCLUDED THE SULFATED TERMINALS AND FAULTY CONTACTORS, WAS COMPLETED TO RESOLVE THE REPORTED THERMAL DAMAGE. THE CABLE GAUGE WAS ALSO REPLACED IN ORDER TO SUPPORT A GREATER LOAD OF CURRENT INTENSITY. IT WAS REPORTED THAT THE MACHINE HAS APPROXIMATELY 5,320 HOURS. IT WAS NOT REPORTED THAT ANY SAMPLES ARE AVAILABLE TO BE RETURNED TO THE MANUFACTURER FOR PHYSICAL EVALUATION. THE MACHINE DATA IS AVAILABLE FOR REVIEW BY THE MANUFACTURER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1023038 AQUABPLUS, HF 3000 SUBSYSTEM, WATER PURIFICATION FIP VIVONIC GMBH

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown