PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM
Report
- Report Number
- 2953144-2010-01598
- Event Type
- Injury
- Date Received
- August 12, 2010
- Date of Event
- June 23, 2010
- Report Date
- July 19, 2010
- Manufacturer
- ABBOTT VASCULAR - REDWOOD CITY
- Product Code
- MGB
- PMA / PMN Number
- P960043
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IA, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4) - PT SELECTION- (B)(4). DEVICE # 1: PART # 12673-03, LOT# 90005-6H INDICATED IS BEING FILED UNDER A SEPARATE MEDWATCH MFR NUMBER. THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT COMPLETE.
DEVICE # 2 ISSUE: NEEDLE-TO-CUFF MISS. TIME OF DEVICE ISSUE: DURING VESSEL CLOSURE. ADVERSE EVENT: FAILURE TO ACHIEVE HEMOSTASIS. IT WAS REPORTED THAT A PHYSICIAN TRAINED IN THE USE OF THE PERCLOSE PROGLIDE DEVICE, ATTEMPTED ARTERIOTOMY CLOSURE OF A MODERATELY CALCIFIED COMMON FEMORAL ARTERY AFTER AN INTERVENTIONAL PROCEDURE. REPORTEDLY, A NEEDLE-TO-CUFF MISS OCCURRED. WHEN THE NEEDLE PLUNGER WAS REMOVED, NO SUTURE WAS PRESET. THE DEVICE WAS REMOVED OVER A GUIDE WIRE AND A SECOND PROGLIDE DEVICE WAS USED WITH THE SAME RESULTS. A THIRD PROGLIDE WAS USED TO ACHIEVE HEMOSTASIS. THERE WERE NO REPORTED PT ADVERSE EFFECTS. NO ADDITIONAL INFO WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM | MGB | ABBOTT VASCULAR - REDWOOD CITY | NA | 90005-6H |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention | DEVICE #1: PART # 12673-03, LOT# 90005-6H |