FDA Adverse Event Injury Summary report: N

PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM

MDR report key: 1800493 · Received August 12, 2010

Report

Report Number
2953144-2010-01603
Event Type
Injury
Date Received
August 12, 2010
Date of Event
June 30, 2010
Report Date
July 19, 2010
Manufacturer
ABBOTT VASCULAR - REDWOOD CITY
Product Code
MGB
PMA / PMN Number
P960043
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT COMPLETE.

Description of Event or Problem · 1

DEVICE ISSUE: NEEDLE-TO-CUFF MISS. TIME OF DEVICE ISSUE: DURING VESSEL CLOSURE. ADVERSE EVENT: FAILURE TO ACHIEVE HEMOSTASIS. IT WAS REPORTED THAT A PHYSICIAN TRAINED IN THE USE OF THE PERCLOSE PROGLIDE DEVICE, ATTEMPTED ARTERIOTOMY CLOSURE OF A MODERATELY CALCIFIED COMMON FEMORAL ARTERY AFTER AN INTERVENTIONAL PROCEDURE. REPORTEDLY, A NEEDLE-TO-CUFF MISS OCCURRED. WHEN THE NEEDLE PLUNGER WAS REMOVED, NO SUTURE WAS PRESENT. THE DEVICE WAS REMOVED OVER A GUIDE WIRE AND A SECOND PROGLIDE DEVICE WAS USED TO ACHIEVE HEMOSTASIS. THERE WERE NO REPORTED PT ADVERSE EFFECTS. NO ADDITIONAL INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM MGB ABBOTT VASCULAR - REDWOOD CITY NA 90005-6H

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention