FDA Adverse Event Malfunction Summary report: N

ECHELON 60

MDR report key: 1800398 · Received August 16, 2010

Report

Report Number
3005075853-2010-04649
Event Type
Malfunction
Date Received
August 16, 2010
Date of Event
June 25, 2010
Report Date
July 6, 2010
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K051002
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). LOCKOUT CARTRIDGE. THE (B)(4) DEVICE WAS RECEIVED FOR ANALYSIS WITH NO VISUAL NON-CONFORMANCES AND WITH AN (B)(4) CARTRIDGE RELOAD PRESENT. THE CARTRIDGE RELOAD WAS RECEIVED PARTIALLY FIRED 1/16. IT IS POSSIBLE THAT WHILE LOADING THE RELOAD, THE CARTRIDGE WAS PUSHED FARTHER BACK THAN THE CARTRIDGE ALIGNMENT STOP WINDOWS RESULTING IN THE KNIFE PUSHING THE ONE PIECE SLED FORWARD, LOCKING THE CARTRIDGE AND PUSHING STAPLES UPWARDS. THE RETURNED CARTRIDGE RELOAD WAS TESTED FOR FUNCTIONALITY BY RESETTING AND RELOADING IT INTO THE DEVICE. THE FUNCTIONAL TEST DEMONSTRATED THAT THE REMAINING STAPLES IN THE CARTRIDGE RELOAD FORMED PROPERLY AND THE INSTRUMENT ACHIEVED ITS COMPLETE 3-STROKE FIRING SEQUENCE WITHOUT ANY DIFFICULTIES. THE DEVICE OPENED AND CLOSED AS INTENDED. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A GASTRIC BYPASS PROCEDURE, THE FIRST FIRING WENT SMOOTH, THE SECOND FIRING STAGNATED. ANOTHER DEVICE WAS USED TO FINISH THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ECHELON 60 STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. UNK G4RX8J

Patients

Seq Age Sex Outcome Treatment
1