T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY
MDR report key: 18003089
·
Received October 25, 2023
Report
- Report Number
- 3013756811-2023-150303
- Event Type
- Malfunction
- Date Received
- October 25, 2023
- Date of Event
- September 19, 2023
- Report Date
- October 25, 2023
- Manufacturer
- TANDEM DIABETES CARE
- Product Code
- QFG
- UDI-DI
- 00389152000077
- PMA / PMN Number
- K201214
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 0
NO PRODUCT WAS RETURNED FOR EVALUATION. SHOULD NEW RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE TANDEM PUMP USER GUIDE INSTRUCTS THE USER TO REMOVE ANY RESIDUAL AIR FROM THE CARTRIDGE. THE CARTRIDGE USER GUIDE CAUTIONS THAT INSULIN SHOULD BE AT ROOM TEMPERATURE BEFORE FILLING A CARTRIDGE.
Description of Event or Problem · 0
IT WAS REPORTED THAT A MINIMUM FILL NOTIFICATION OCCURRED AFTER THE USER FILLED THE CARTRIDGE WITH 300 UNITS OF INSULIN DURING THE LOAD SEQUENCE. REPORTEDLY, THE CUSTOMER DID NOT PERFORM THE PROPER AIR REMOVAL TECHNIQUES AND LOADED COLD INSULIN INTO THE CARTRIDGE. A NEW CARTRIDGE WAS LOADED TO RESOLVE THE ISSUE. CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 200-212 MG/DL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1433887 | T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY | ALTERNATE CONTROLLER ENABLED INFUSION PUMP | QFG | TANDEM DIABETES CARE | 1002717 | 00389152000077 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 16 YR | Male |