33CM PKS OMNI INSTRUMENT
Report
- Report Number
- 2183680-2010-00040
- Event Type
- Malfunction
- Date Received
- August 10, 2010
- Date of Event
- July 20, 2010
- Report Date
- August 10, 2010
- Manufacturer
- GYRUS MEDICAL, INC.
- Product Code
- GEI
- PMA / PMN Number
- K081766
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
THE COMPLAINT IS CONFIRMED, THE HUB WAS FRACTURED UNDER THE SHRINK TUBING OF THE SHAFT. THE SHRINK TUBING WAS REMOVED TO EXPOSE THE HUB. THE FRACTURE CONSISTED OF TWO LARGE PIECES AND SEVERAL SMALL FRAGMENTS. BREAKAGE OCCURRED AT THE DISTAL END OF THE SHAFT. DAMAGE TO THE ELECTRICAL WIRES WAS FOUND AND APPEARS TO BE RELATED TO THE FRACTURE OF THE HUB. SEVERAL SMALL FRAGMENTS WERE FOUND IMBEDDED IN THE SHAFT SHRINK TUBING. WHEN COMPARED TO AN INTACT HUB, WE CANNOT VERIFY THAT ALL OF THE FRAGMENTS ARE ACCOUNTED FOR. THERE WAS NO REPORTED PT HARM AND THE HUB FRACTURE OCCURRED IN A LOCATION UNDER THE SHRINK TUBING WHICH MAY HAVE PREVENTED THE FRAGMENTS FROM ESCAPING INTO THE BODY. THE EXTENT OF THE HUB FRACTURE COULD NOT BE DETECTED UNLESS THE SHRINK TUBING WAS REMOVED. CARDIFF, UK'S R&D TEAM HAS BEEN WORKING ON A HUB REDESIGN WHICH WILL ADDRESS THIS HUB FRACTURING ISSUE. THIS REDESIGN HAS NOT YET BEEN IMPLEMENTED TO DATE. HUB FRACTURES HAVE BEEN ATTRIBUTED TO EXCESSIVE FORCE BEING APPLIED TO THE DEVICE BEYOND ITS DESIGNED SPECIFICATIONS.
DURING A TOTAL LAPAROSCOPIC HYSTERECTOMY, THE HUB ON THE PKS OMNI INSTRUMENT FRACTURED AND THE JAWS WOULD NOT OPERATE. THE DEVICE WAS REMOVED FROM THE OPERATING SITE AND ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH NO PT HARM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 33CM PKS OMNI INSTRUMENT | PKS OMNI INSTRUMENT, 33CM | GEI | GYRUS MEDICAL, INC. | 970010PC | 143956MB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |