FDA Adverse Event Malfunction Summary report: N

FLUOROSTAR

MDR report key: 1800058 · Received August 11, 2010

Report

Report Number
9680959-2010-00267
Event Type
Malfunction
Date Received
August 11, 2010
Date of Event
June 14, 2010
Report Date
August 11, 2010
Manufacturer
GE OEC MEDICAL SYSTEMS GMBH
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE FAILURE COULD NOT BE DUPLICATED. THE COLLIMATOR WAS REPLACED AND CALIBRATED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE 7900 SYSTEM HAD A COLLIMATOR IRIS POTENTIOMETER ERROR MESSAGE DURING A CASE. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FLUOROSTAR FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS GMBH 7900

Patients

Seq Age Sex Outcome Treatment
1