FDA Adverse Event Malfunction Summary report: N

LEGEND DISSECTING TOOL

MDR report key: 1800051 · Received August 11, 2010

Report

Report Number
1625507-2010-00044
Event Type
Malfunction
Date Received
August 11, 2010
Date of Event
July 9, 2010
Report Date
July 13, 2010
Manufacturer
MEDTRONIC POWERED SURGICAL SOLUTIONS
Product Code
HBE
PMA / PMN Number
K020069
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SN
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

REPORT CONFIRMED. EVALUATION DETERMINED THAT THE FRACTURE PLANE WAS PERPENDICULAR TO THE STEM AXIS. THE FRACTURE SURFACES WERE ALSO "SMEARED" INDICATING THAT THE TWO SURFACES WERE RUBBED TOGETHER AFTER THE FRACTURE OCCURRED. ON FOLLOW-UP, IT WAS CONFIRMED THAT THERE WAS NO PT IMPACT. THE USER MANUAL CONTAINED THE FOLLOWING "DO NOT USE EXCESSIVE PRESSURE, SUCH AS BENDING OR PRYING, ON ATTACHMENTS OR DISSECTING TOOLS. THIS MAY CAUSE TOOL TO BEND OR BREAK AND CAUSE INJURY TO PT, OPERATOR AND/OR OPERATING ROOM STAFF."

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SURGEON WAS USING A LEGEND MOTOR, AN AF02 ATTACHMENT AND AN F28TA23 TOOL. TOOL BROKE AS HE WAS IN THE MIDDLE OF CRANIOTOMY CUTTING THE BONE FLAP. NO PT IMPACT WAS REPORTED. ON FOLLOW-UP, IT WAS NOTED THAT THE DETACHED PORTION WAS EASILY RETRIEVED AND IT WAS CONFIRMED THAT THERE WAS NO PT IMPACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LEGEND DISSECTING TOOL HBE-DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, HBE MEDTRONIC POWERED SURGICAL SOLUTIONS NA 0002019880

Patients

Seq Age Sex Outcome Treatment
1 46 YR