LEGEND DISSECTING TOOL
Report
- Report Number
- 1625507-2010-00044
- Event Type
- Malfunction
- Date Received
- August 11, 2010
- Date of Event
- July 9, 2010
- Report Date
- July 13, 2010
- Manufacturer
- MEDTRONIC POWERED SURGICAL SOLUTIONS
- Product Code
- HBE
- PMA / PMN Number
- K020069
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SN
- Reporter Occupation
- NURSE
Narratives
REPORT CONFIRMED. EVALUATION DETERMINED THAT THE FRACTURE PLANE WAS PERPENDICULAR TO THE STEM AXIS. THE FRACTURE SURFACES WERE ALSO "SMEARED" INDICATING THAT THE TWO SURFACES WERE RUBBED TOGETHER AFTER THE FRACTURE OCCURRED. ON FOLLOW-UP, IT WAS CONFIRMED THAT THERE WAS NO PT IMPACT. THE USER MANUAL CONTAINED THE FOLLOWING "DO NOT USE EXCESSIVE PRESSURE, SUCH AS BENDING OR PRYING, ON ATTACHMENTS OR DISSECTING TOOLS. THIS MAY CAUSE TOOL TO BEND OR BREAK AND CAUSE INJURY TO PT, OPERATOR AND/OR OPERATING ROOM STAFF."
IT WAS REPORTED THAT THE SURGEON WAS USING A LEGEND MOTOR, AN AF02 ATTACHMENT AND AN F28TA23 TOOL. TOOL BROKE AS HE WAS IN THE MIDDLE OF CRANIOTOMY CUTTING THE BONE FLAP. NO PT IMPACT WAS REPORTED. ON FOLLOW-UP, IT WAS NOTED THAT THE DETACHED PORTION WAS EASILY RETRIEVED AND IT WAS CONFIRMED THAT THERE WAS NO PT IMPACT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LEGEND DISSECTING TOOL | HBE-DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, | HBE | MEDTRONIC POWERED SURGICAL SOLUTIONS | NA | 0002019880 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 46 YR |