AGILITRAC .035 PERIPHERAL DILATATION CATHETER
Report
- Report Number
- 3004742046-2010-00355
- Event Type
- Malfunction
- Date Received
- August 12, 2010
- Date of Event
- July 12, 2010
- Report Date
- July 19, 2010
- Manufacturer
- ABBOTT VASCULAR-VASCULAR SOLUTIONS
- Product Code
- LIT
- PMA / PMN Number
- K081417
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). THE AGILITRAC .035 PERIPHERAL DILATATION CATHETER, PART # 1010005-40, LOT # 9100851 INDICATED IS BEING FILED UNDER A SEPARATE MEDWATCH MFR NUMBER. EVALUATION SUMMARY: THE BALLOON CATHETER WAS RETURNED WITH BLOOD ON THE BALLOON AND CONTRAST IN THE BALLOON. THE BALLOON WAS LOOSELY FOLDED. THERE WAS A LONGITUDINAL RUPTURE 2MM DISTAL TO THE PROXIMAL MARKER FOR A LENGTH OF 2CM. THERE WERE LONGITUDINAL SCRATCHES AT THE PROXIMAL END OF THE RUPTURE, 2MM IN LENGTH. THERE WAS NO OTHER DAMAGE NOTED TO THE BALLOON CATHETER. BALLOON RUPTURES CAN OCCUR AS A RESULT OF MANUFACTURING OR FROM USE OF THE PRODUCT. DURING USE THERE CAN BE AN INTERACTION WITH ANATOMY AND/OR ACCESSORY DEVICES, WHICH CAN DAMAGE OR WEAKEN THE BALLOON MATERIAL AND LEAD TO RUPTURE DURING INFLATION. THERE WAS NO REPORT OF ANY LEAK IN THE PRODUCT NOTED DURING PREPARATION FOR USE, WHICH WOULD SUGGEST THAT THE BALLOON WAS NOT DAMAGED PRIOR TO USE. TO ENSURE THIS TYPE OF DAMAGE IS NOT A RESULT OF MANUFACTURING, ALL BALLOON CATHETERS ARE LEAK TESTED ON-LINE AND A SAMPLING OF UNITS FROM ALL CATHETER LOTS ARE RUPTURE TESTED PRIOR TO RELEASE FOR USE IN THE FINISHED GOODS LOT. IN THIS CASE, THE SCRATCHES NOTED PROXIMAL TO THE RUPTURE SUGGEST THAT THE BALLOON INTERACTED WITH THE ANATOMY AND/OR ACCESSORY DEVICES. ADDITIONALLY, THE RUPTURE AND BALLOON DAMAGE WAS ALSO NOTED ON THE SECOND RETURNED AGILITRAC USED IN THIS PROCEDURE, FURTHER SUGGESTING THAT THE RUPTURES ARE LIKELY RELATED TO INTERACTION WITH THE ANATOMY AND/OR ACCESSORY DEVICES. A REVIEW OF THE FINISHED DEVICE LOT HISTORY RECORDS DID NOT REVEAL ANY NONCONFORMING MATERIAL RECORDS FOR THIS LOT, AND THERE HAVE BEEN NO OTHER INCIDENTS REPORTED FOR THIS LOT. ADDITIONALLY, LOT RELEASE TESTING RESULTS WERE REVIEWED AND ALL SAMPLES MET ALL MANUFACTURING CRITERIA. IN THIS CASE, THE BALLOON RUPTURE APPEARS TO BE RELATED TO CASE CIRCUMSTANCES AND THERE IS NO INDICATION OF A PRODUCT RELATED DEFICIENCY.
DEVICE ISSUE: BALLOON RUPTURE. TIME OF DEVICE ISSUE: DURING PROCEDURE. ADVERSE EVENT: NONE. IT WAS REPORTED THAT DURING DILATATION IN AN UNSPECIFIED VESSEL, THE AGILITRAC DILATATION CATHETER BALLOON RUPTURED BELOW 14 ATMOSPHERES. A SECOND AGILITRAC DILATATION CATHETER BALLOON ALSO RUPTURED BELOW 14 ATMOSPHERES. THERE WAS NO ADVERSE PATIENT EFFECT REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AGILITRAC .035 PERIPHERAL DILATATION CATHETER | LIT | ABBOTT VASCULAR-VASCULAR SOLUTIONS | NA | 9063051 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 60 YR | PART # 1010005-40, LOT # 9100851| AGILITRAC .035 PERIPHERAL DILATATION CATHETER |