FDA Adverse Event Malfunction Summary report: N

TALENT THORACIC STENT GRAFT SYSTEM - XCELERANT

MDR report key: 1800029 · Received August 11, 2010

Report

Report Number
2953200-2010-01532
Event Type
Malfunction
Date Received
August 11, 2010
Date of Event
July 14, 2010
Report Date
July 14, 2010
Manufacturer
MEDTRONIC CARDIOVASCULAR GALWAY
Product Code
MIH
PMA / PMN Number
P070007
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). RESULTS: AORTIC ARCH WAS SEVERELY ANGULATED, WEAKENED VESSEL WALL. CONCLUSIONS: AORTIC ARCH WAS SEVERELY ANGULATED, WEAKENED VESSEL WALL.

Description of Event or Problem · 1

A TALENT THORACIC STENT GRAFT SYSTEM WAS IMPLANTED IN THE PATIENT FOR TREATMENT TRANSECTION IN THE THORACIC AORTA. THE PATIENT PRESENTED EMERGENTLY WITH CHEST PAIN AND A HOARSE VOICE. THE PATIENT HAD BEEN IN A MOTOR VEHICLE ACCIDENT ABOUT 6 MONTHS AGO. THE REVIEW OF THE CT DEMONSTRATES THERE WAS A TRANSECTION OF THE AORTA. THE PHYSICIAN BELIEVES THE TRANSECTION OF THE AORTA WAS MOST LIKELY DUE TO THE MOTOR VEHICLE ACCIDENT, WEAKENING THE VESSEL WALL. THE AORTIC ARCH WAS MODERATELY ANGULATED. IT WAS REPORTED THAT THE PROXIMAL MAIN DEVICE WAS PLACED AT THE LEFT SUBCLAVIAN AND THE APEX OF THE STENT GRAFT MISALIGNED DURING DEPLOYMENT. THE PHYSICIAN WAS ABLE TO CORRECT THE MISALIGNMENT OF THE STENT GRAFT, BY PULLING THE STENT GRAFT DOWN, HOWEVER, THIS RESULTED IN INACCURATE DELIVERY OF THE DEVICE. (MFR 2953200-2010-01531) THE PHYSICIAN ELECTED TO USE A SECOND DEVICE AND THE SECOND DEVICE ALSO MISALIGNED HOWEVER, WAS UNABLE TO CORRECT THE MISALIGNED DEPLOYMENT. (MFR 2953200-2010-01532). THERE WAS A PROXIMAL SEAL AND THE DECISION WAS MADE NOT TO FURTHER INTERVENE AND TO MONITOR THE PATIENT. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED, AND THE PATIENT IS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TALENT THORACIC STENT GRAFT SYSTEM - XCELERANT MIH MEDTRONIC CARDIOVASCULAR GALWAY NA V00428547

Patients

Seq Age Sex Outcome Treatment
1 79 YR