FDA Adverse Event Malfunction Summary report: N

BD BACTEC¿ LYTIC/10 ANAEROBIC/F CULTURE VIALS (PLASTIC)

MDR report key: 18000150 · Received October 24, 2023

Report

Report Number
3008352382-2023-00282
Event Type
Malfunction
Date Received
October 24, 2023
Date of Event
September 13, 2023
Report Date
December 7, 2023
Manufacturer
BECTON DICKINSON CARIBE LTD.
Product Code
MDB
UDI-DI
00382904420215
PMA / PMN Number
K123903
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.6 INVESTIGATION SUMMARY CATALOG: 442021 BATCH NO.: 3109880, 3130606 & 3138057. CUSTOMER REPORTED A MOLECULAR FALSE POSITIVE RESULT. NEITHER PHOTOS NOR RETURNED GOOD SAMPLES WERE RECEIVED. BD WAS UNABLE TO REPRODUCE THE CUSTOMER¿S EXPERIENCE WITH THE BACTEC PRODUCT BASED ON OUR INTERNAL PROCEDURES AND THE INTENDED USE OF THE PRODUCT. RETENTION SAMPLES WERE VISUALLY INSPECTED, TESTED FOR VIABLE CONTAMINATION BY SUB-CULTURING ON TSA, CHOCOLATE, SABOURAUD AND SCHAEDLER AGARS PLATES, VOLTAGE OUTPUT AND GRAM STAIN. ALL RESULTS WERE SATISFACTORY. BATCH HISTORY RECORD REVIEW DID NOT IDENTIFY ANY EVIDENCE FOR WHICH THE CUSTOMER SUBMITTED THE COMPLAINT. A COMPLAINT HISTORY REVIEW WAS CONDUCTED, AND BATCH HAS BEEN PREVIOUSLY INVESTIGATED FOR THE REPORTED DEFECT. COMPLAINT IS UNCONFIRMED BASED ON RETENTION SAMPLES AND BATCH HISTORY RECORD REVIEW RESULTS. NO CORRECTIVE ACTIONS WERE REQUIRED. A CROSS FUNCTIONAL TEAM CONTINUALLY MONITORS ALL PRODUCT COMPLAINTS FOR TRENDS AND DETERMINES IF ANY ADDITIONAL ACTIONS ARE NECESSARY BEYOND THE CURRENT INVESTIGATIONAL PROCESS.

Additional Manufacturer Narrative · 0

H.3. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING BD BACTEC¿ LYTIC/10 ANAEROBIC/F CULTURE VIALS (PLASTIC) MOLECULAR FALSE POSITIVE FOR C. TROPICALIS OCCURRED. CONFIRMATORY TESTING ONLY GREW PAENIBACILLUS . THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER STATES THEY HAVE MULTIPLE MOLECULAR FP FOR C. TROPICALIS WHAT CONFIRMATORY TESTING WAS PERFORMED THAT DETERMINED THE RESULTS WERE ERRONEOUS? SUB-CULTURE WAS NEGATIVE FOR C. TROPICALIS CULTURE RESULTS: PAENIBACILLUS.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING BD BACTEC¿ LYTIC/10 ANAEROBIC/F CULTURE VIALS (PLASTIC) MOLECULAR FALSE POSITIVE FOR C. TROPICALIS OCCURRED. CONFIRMATORY TESTING ONLY GREW PAENIBACILLUS . THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER STATES THEY HAVE MULTIPLE MOLECULAR FP FOR C. TROPICALIS. WHAT CONFIRMATORY TESTING WAS PERFORMED THAT DETERMINED THE RESULTS WERE ERRONEOUS? SUB-CULTURE WAS NEGATIVE FOR C. TROPICALIS. CULTURE RESULTS: PAENIBACILLUS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1579456 BD BACTEC¿ LYTIC/10 ANAEROBIC/F CULTURE VIALS (PLASTIC) SYSTEM, BLOOD CULTURING MDB BECTON DICKINSON CARIBE LTD. 3130606 00382904420215

Patients

Seq Age Sex Outcome Treatment
1 Unknown