FDA Adverse Event Malfunction Summary report: N

SINGLE-SITE

MDR report key: 17999895 · Received October 24, 2023

Report

Report Number
2955842-2023-19568
Event Type
Malfunction
Date Received
October 24, 2023
Date of Event
September 26, 2023
Report Date
September 26, 2023
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
00886874113585
PMA / PMN Number
K152448
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

AN INVESTIGATION WAS COMPLETED TO DETERMINE THE CAUSE OF THIS REPORTED EVENT. INTUITIVE SURGICAL, INC. (ISI) RECEIVED THE DA VINCI PRODUCT TO PERFORM FAILURE ANALYSIS. FAILURE ANALYSIS COULD NOT CONFIRM NOR REPLICATE THE REPORTED ISSUE. THE MEDIUM-LARGE CLIP APPLIER WAS DRIVEN ON AN IN-HOUSE SYSTEM AND PASSED THE RECOGNITION AND ENGAGEMENT TESTS. THE INSTRUMENT MOVED INTUITIVELY WITH THE FULL RANGE OF MOTION IN ALL DIRECTIONS. THE GRIPS OPENED AND CLOSED PROPERLY. THERE WAS NO ISSUE WITH THE CLIPS ATTACHING TO THE INSTRUMENT OR CLAMPING. THE INSTRUMENT WAS FULLY FUNCTIONAL. .

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED CHOLECYSTECTOMY, THE MEDIUM-LARGE CLIP APPLIER INSTRUMENT WOULD NOT RELEASE THE CLIP WHEN APPLYING IT ON A CYSTIC ARTERY. THE MEDIUM-LARGE CLIP APPLIER WOULD NOT ENGAGE ON THE UNIVERSAL SURGICAL MANIPULATOR (USM) 1 OF THE PATIENT SIDE CART (PSC) AFTER IT WAS REMOVED. AN INTUITIVE SURGICAL, INC. (ISI) TECHNICAL SUPPORT ENGINEER (TSE) REVIEWED THE SYSTEM LOGS AND CONFIRMED ERROR 22020 ON USM1. THE TSE ASKED THE CUSTOMER TO RESEAT THE DRAPE ON USM1 AND TRIED ANOTHER INSTRUMENT. THE CUSTOMER WAS ABLE TO INSTALL A DIFFERENT INSTRUMENT SUCCESSFULLY ON USM1. THE PROCEDURE WAS COMPLETED WITH NO PATIENT HARM. INTUITIVE SURGICAL, INC. (ISI) MADE MULTIPLE FOLLOW-UP ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION. HOWEVER, NO FURTHER DETAILS HAVE BEEN RECEIVED AS OF THE DATE OF THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1579432 SINGLE-SITE CLIP APPLIER, MEDIUM-LARGE NAY INTUITIVE SURGICAL, INC 478053-04 K10230323 0054 00886874113585

Patients

Seq Age Sex Outcome Treatment
1 Unknown DA VINCI INSTRUMENTS AND ACCESSORIES